2.5 Reduced Pedal Pumps
Chapter 2: Troubleshooting Procedures
2.5 Reduced Pedal Pumps
1.
The stretcher rises to the high position after approximately 28 presses of
the Hilow Up pedal.
Yes
No
↓
→
Remove air from the hydraulic system (refer to procedure 4.29).
If this solves the problem, go to “Final Actions” on page 2-11.
Otherwise, continue to step 2.
2.
Lower the stretcher, and press the Hilow Up pedal approximately 28 times
to raise the stretcher to the high position. Both the head and foot ends of the
stretcher raise evenly.
Yes
No
↓
→
If the head end does not rise, replace the hydraulic cylinder at the
head end of the stretcher. If the foot end does not rise, replace the
hydraulic cylinder at the foot end of the stretcher (refer to
procedure 4.29). If this solves the problem, go to “Final
Actions” on page 2-11. Otherwise, call Technical Support.
3.
Go to “Final Actions” on page 2-11.
Page 2 - 16
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)