4.24 OB/GYN Stretcher—Foot Support Mechlok® Device
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.24 OB/GYN Stretcher—Foot Support Mechlok®
1
Device
Tools required:
Ratchet
3/16" hex head bit
1/8" hex head bit
Screwdriver
Retaining ring removal/installation tool
Removal
1. Remove the mattress assembly from the stretcher.
WARNING:
Make sure the stretcher is in the full flat position, and the brake/steer
pedal is in the brake position. Failure to do so could cause personal
injury or equipment damage.
2.
Put the stretcher in the full flat position.
3.
Remove the foot section from the stretcher.
4.
Remove the two screws (J) that attach the foot section guide (K) to the
upper frame carriage weldment (L) (see figure 4-35 on page 4-71).
Figure 4-35. Foot Section Guide
4
1. Mechlok® is a registered trademark of P. L. Porter Company.
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Page 4 - 71
Service Manual (144386 REV 2)