The MetaNeb® System User Manual (174432 REV 7)
21
Cleaning
WARNING:
Do not use phenolic, alcohol, quaternary ammonium chloride, or bleach
solutions to clean the patient circuit. Use these solutions to clean the
controller only. Injury or equipment damage can occur.
The Metaneb® Controller has been tested for compatibility with the
following cleaning and disinfecting solutions:
Cleaning the Controller and Stand
Clean the controller and stand between patients, when visibly soiled, or
according to facility protocols. Do not spray the cleaner on to the
controller. We recommend that you clean the controller and stand with a
soft cotton cleaning pad that is moistened with the cleaner.
Do not use excessive liquid or harsh cleansers. Do
not
immerse the
controller in water or let liquids enter the controller.
Clean the controller and stand as follows:
a.
Make sure the air supply hose is connected to the controller.
b.
Make sure a circuit is connected.
c.
Wipe down the controller or stand with the moistened cloth.
d.
After you clean the controller or stand, make sure it is dry before
you use it.
Cleaning the Circuit
Chemical Class
Active Ingredient
Phenolic Ortho-Phenylphenol
Ortho-Benzyl-para-Chlorophenol
Alcohol Isopropyl
alcohol
Quaternary
ammonium chloride
Didecyl dimethyl ammonium chloride
Alkyl dimethyl benzyl ammonium
chloride
Bleach
Sodium hypochlorite
Summary of Contents for MetaNeb System
Page 1: ...The MetaNeb System User Manual Product No PMN4 174432 REV 7...
Page 2: ......
Page 35: ...The MetaNeb System User Manual 174432 REV 7 29 Specifications Controller Dimensions...
Page 48: ...42 The MetaNeb System User Manual 174432 REV 7 Particle Specifications NOTES...
Page 49: ......
Page 50: ......