Hill-Rom Liko Viking S Instruction Manual Download Page 7

7

www.liko.com

Viking S / Viking XS  •  7EN138104-01 

 Never move the lift by pulling on the actuator!

Electrical Emergency Lowering

Lower the lift arm by pressing the marking 

 

for ”down” (     ) on the control unit.

Mechanical Emergency Lowering

Turn the red emergency lowering cylinder 

 

in the direction of the arrows.

Emergency Stop

To activate the emergency stop: 

Press the red 

button.

To reset the emergency stop: 

Turn the red button 

in the direction of the arrows.

Locking the Wheels

The rear wheels can be locked to prevent rotating and 

turning. To lock the wheels, push down the lock pedal 

 

with your foot. To unlock the wheels, lift up the lock 

 

pedal with your foot.
When lifting, the wheels should be unlocked so that the lift 

can be moved to the patient’s centre of gravity. The wheels 

should be locked however, if there is a risk of the lift rolling 

into the patient, for instance, when lifting from the floor.

     Locked wheels during lifting can increase the risk

of tipping.

Installation of Latches

After installation, check that the latch locks 

 

and runs freely in the sling bar hook.

Unlock
Lock

Summary of Contents for Liko Viking S

Page 1: ...contents of the instruction guides The equipment should only be used by trained personnel Please contact your Liko Hill Rom representative in the event of any uncertainties or questions Product Descri...

Page 2: ...Viking XS have been tested by an accredited testing institute and fulfil the requirements specified in the Medical Device Directive 93 42 EEC for Class 1 products Viking S Viking XS complies with the...

Page 3: ...ulated lead acid gel type batteries New batteries are provided by the supplier Battery charger CH01 In 100 240 V AC 50 60 Hz max 250 mA Out 24 V max 500 mA Lift motor 24 V permanent magnetic motor wit...

Page 4: ...rence measurement according to Standard EN ISO 10535 2006 Lateral view B B1 B2 N O Max reach of lift arm L M F F1 600 mm 23 6 in Ref 1 B3 Amax E Amin 700 mm 27 5 in Top view C D2 D D1 Ref 2 A max A mi...

Page 5: ...ry charger lift mast with lift arm motor for lift arm sling bar and control unit base lever for base width adjustment hand control with cable 2 Connect the cables as follows Cable from hand control to...

Page 6: ...arrows 5 Before the first use the lift s battery should be charged for at least 12 hours For detailed instructions see Charging the Batteries p 8 3 1 2 Hand Control Indicator light A charge the batter...

Page 7: ...the arrows Locking the Wheels The rear wheels can be locked to prevent rotating and turning To lock the wheels push down the lock pedal with your foot To unlock the wheels lift up the lock pedal with...

Page 8: ...obtain maximum life time it is important to charge the batteries regularly We recommend charging the batteries after each use of the lift or every night If the lift is not used every day we recommend...

Page 9: ...3156079 incl bag Max 300 kg 660 lbs Sling Cross bar 450 Prod No 3156021 Max 300 kg 660 lbs Can be used to a limited extent with Viking XS Sling Cross bar 670 Prod No 3156018 Max 300 kg 660 lbs Can be...

Page 10: ...No 3156502 Adapter 12 mm Prod No 2016504 LikoScale 350 Prod No 3156228 Quick Release Hook Liko s Quick Release Hooks are a system for quick change of lifting accessories on Liko s mobile and stationar...

Page 11: ...nected to a wall socket If the problem remains please contact Liko Hill Rom 1 Check that the emergency stop button is not activated 2 Check that the wall socket is energized 3 Check that the charging...

Page 12: ...sing original Liko spare parts Service Agreement Liko offers the opportunity to enter into service contracts for the maintenance and regular inspection of your Liko product Expected Life Time The prod...

Reviews: