Hill-Rom careselections Synergy Air Elite User Manual Download Page 9

Synergy® Air Elite Low Airloss Alternating Therapy System User Manual

 3

(139122 REV 1)

Power Failure

—For 

additional safety, the 

Power Failure

 alarm and 

indicator come on when 
the power to the control unit fails or the main 

Power

 switch is turned 

off.

Alarm Reset

—Press to turn the 

Power Failure

 alarm off.

Specifications

Feature

Dimension

Control unit length

16.5" (42 cm) 

Control unit width

6.7" (17 cm)

Control unit height

10.6" (27 cm)

Control unit weight

13.2 lb (6.0 kg)

Control unit phase time

5 minutes to 95 minutes

Control unit maximum flow 
rate

1200 lpm

Control unit rated voltage

110 V to 120 V

Control unit maximum 
current

5 A

Control unit rated frequency

60 Hz

Control unit protection type

Class II, Type BF

Control unit mode of 
operation

Continuous

Control unit safety standard

Conformity European (CE) and 
Underwriters Laboratories Inc.®

a

 (UL)

Mattress width

36" (91 cm), 39" (99 cm), or 48" 
(122 cm)

Mattress length

79.9" (203 cm)

Mattress height

9.4" (24 cm)

Mattress weight

33.1 lb (15.0 kg)

Air cell material

Navy blue, nylon-woven fabric with 
polyurethane back

Coverlet type

Fully-quilted zipper cover with a 
removable, 2", convoluted foam base

Coverlet material

Navy blue, woven fabric with 
polyurethane coat finish

Summary of Contents for careselections Synergy Air Elite

Page 1: ...USER MANUAL Synergy Air Elite Low Airloss Alternating Therapy System From Hill Rom 139122 REV 1...

Page 2: ......

Page 3: ...l Rom is a registered trademark of Hill Rom Services Inc Synergy is a registered trademark of Mediq PRN Life Support Services Inc Underwriters Laboratories Inc is a registered trademark of Underwriter...

Page 4: ...ii Synergy Air Elite Low Airloss Alternating Therapy System User Manual 139122 REV 1 Revision Pages Affected Date 1 December 2005...

Page 5: ...ating Therapy Mode 8 Static Mode 9 Auto Firm Mode 9 Power Failure Alarm 9 Cleaning 10 General Cleaning 10 Control Unit 10 Mattress 11 Steam Cleaning 11 Cleaning Hard to Clean Spots 12 Disinfecting 12...

Page 6: ...iv Synergy Air Elite Low Airloss Alternating Therapy System User Manual 139122 REV 1...

Page 7: ...oldface text emphasizes a word or phrase NOTE sets apart special data or important instruction clarification WARNING or CAUTION A WARNING identifies situations or actions that may have an effect on pa...

Page 8: ...ntrol Soft Firm Press the Soft or Firm control to adjust the Comfort Control setting shown on the eight increment scale Cycle Timer Press the Cycle Timer control to adjust the Alternating Therapy cycl...

Page 9: ...95 minutes Control unit maximum flow rate 1200 lpm Control unit rated voltage 110 V to 120 V Control unit maximum current 5 A Control unit rated frequency 60 Hz Control unit protection type Class II...

Page 10: ...the raised position Failure to do either of these could result in serious injury or death NOTE Siderails are intended to be a reminder to the patient of the unit s edges not a patient restraining devi...

Page 11: ...atient injury could occur WARNING Before you plug the control unit into an applicable power source carefully read the instructions Failure to do so could cause patient injury personal injury or equipm...

Page 12: ...drop the control unit Equipment damage could occur CAUTION Do not store the Synergy Air Elite Low Airloss Alternating Therapy System in direct sunlight or extreme cold conditions Equipment damage coul...

Page 13: ...Press the Auto Firm control The Auto Firm indicator comes on and the mattress starts to inflate 12 Let the mattress fully inflate NOTE In Auto Firm mode the mattress should fully inflate within 2 to...

Page 14: ...ry third air cell fully inflates 2 Air cells 2 5 8 11 14 17 deflate 3 Every next third air cell fully inflates 4 Air cells 3 6 9 12 15 18 deflate 5 Every final third air cell fully inflates To start t...

Page 15: ...ells and the foam pad in the mattress assembly Make sure at least 2 5 cm exists between the patient and the top of the foam pad Auto Firm Mode Auto Firm mode quickly inflates the air cells to their ma...

Page 16: ...e Personal injury or equipment damage could occur CAUTION Do not use harsh cleansers solvents or detergents Equipment damage could occur General Cleaning We recommend that you clean the unit with dete...

Page 17: ...h the facility s procedures for contaminated waste b Replace with a clean top coverlet or bottom encasement Wash Industrial laundry a Break wash cycle in cold water for 10 minutes b Main wash cycle I...

Page 18: ...ied on the manufacturer s label Maintenance WARNING Only Hill Rom authorized persons should service the Synergy Air Elite Low Airloss Alternating Therapy System Service by unauthorized persons could c...

Page 19: ...thers with Normal Air Pressure 1 Put the unit in Static mode Unzip the top coverlet from the bottom encasement and examine the air tubes between the air cells and the main air supply line for kinks Ad...

Page 20: ...torage and Handling CAUTION Do not store the Synergy Air Elite Low Airloss Alternating Therapy System in direct sunlight or extreme cold conditions Equipment damage could occur Feature Specification S...

Page 21: ......

Page 22: ...1 0 347 32 35 00 Nordic Region Hill Rom AB Ekbacksv gen 20 168 69 Bromma Sweden Tel 46 0 8 564 353 60 Fax 46 0 8 564 353 61 se marketing hill rom com sterreich Hill Rom Austria GmbH Bueropark Donau Ha...

Reviews: