2.9 IV Pole Malfunction
Chapter 2: Troubleshooting Procedures
CarePorter™ Cart Service Manual (man239)
Page 2 - 19
2
2.9
IV Pole Malfunction
1. Raise the IV pole. The IV pole (A) retains its higher position and releases
when the IV pole release (B) is pulled downward (see figure 2-2 on page
2-19).
Yes
No
↓
→
Figure 2-2. IV Pole
2. Go to “Final Actions” on page 2-5.
3. Another part of the CarePorter™ Cart or an accessory interferes with the
vertical function of the IV pole.
Yes
No
↓
→
Inspect the IV pole release for damage. If necessary, replace the
defective IV pole/arm assembly (refer to procedure 4.16). If this
solves the problem, go to “Final Actions” on page 2-5.
Otherwise, call Hill-Rom Technical Support at (800) 445-3720.
4. Take appropriate actions to eliminate the interference. This solves the
problem
m239_082