2.8 IV Pole/Arm Does Not Rotate
Chapter 2: Troubleshooting Procedures
Page 2 - 18
CarePorter™ Cart Service Manual (man239)
2.8
IV Pole/Arm Does Not Rotate
1. The IV pole/arm rotates without excessive force.
Yes
No
↓
→
2. Go to “Final Actions” on page 2-5.
3. Another part of the CarePorter™ Cart interferes with the IV pole/arm
assembly.
Yes
No
↓
→
Inspect the adjustment of the IV pole/arm assembly (refer to
procedure 4.16). If this solves the problem, go to “Final Actions”
on page 2-5. Otherwise, go to step 5.
4. Take appropriate actions to eliminate the interference. This solves the
problem
Yes
No
↓
→
Call Hill-Rom Technical Support at (800) 445-3720.
5. Adjust the IV pole/arm clutch assemblies (refer to procedure 4.16). This
solves the problem.
Yes
No
↓
→
Call Hill-Rom Technical Support at (800) 445-3720.