Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The CareAssist™ Bed model P1170 is intended for use in the electromagnetic environment specified below. The customer or the user of
the model P1170 should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment—
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6kV Contact
± 8kV Air
± 6kV Contact
± 8kV Air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Radiated RF
IEC 61000-4-3
3 Vrms
80 MHz to 2.5 GHz
3 Vrms
80 MHz to 2.5 GHz
Portable and mobile RF communications
equipment should not be used at close dis-
tances to the P1170 bed. (See Note 2)
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2kV on Power Supply Lines
± 1kV on Input/ Output Lines
± 2kV on Power Supply Lines
± 1kV on Input/ Output Lines
Mains power quality should be that of a
typical commercial or hospital environ-
ment
Surge
IEC 61000-4-5
± 1kV Differential Mode
(line-line)
± 2kV Common Mode
(Line-Ground)
± 1kV Differential Mode
(line-line)
± 2kV Common Mode
(Line-Ground)
Mains power quality should be that of a
typical commercial or hospital environ-
ment
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF communications
equipment (cell phones) should not be
used at close distances to the P1170 bed.
(See Note 2)
Power Frequency
Magnetic Fields
IEC 61000-4-8
3 A/m
3 A/m
The power frequency magnetic field
should be measured in the intended instal-
lation location to assure it is sufficiently
low.
Voltage Dips, Short
Interrupts, & Varia-
tions On Power Sup-
ply Lines
IEC 61000-4-11
< 5% U
T
(95% dip in U
T
for 0.5 cycles )
< 40% U
T
(60% dip in U
T
for 5 cycles )
< 70% U
T
(30% dip in U
T
for 25 cycles )
< 5% U
T
(95% dip in U
T
for 5 seconds )
(See Note 1)
< 5% U
T
(95% dip in U
T
for 0.5 cycles )
< 40% U
T
(60% dip in U
T
for 5 cycles )
< 70% U
T
(30% dip in U
T
for 25 cycles )
< 5% U
T
(95% dip in U
T
for 5 seconds )
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If operation is required during an
extended power outage or interruption, the
model P1170 should be switched to oper-
ate from the backup battery.
Note 1: U
T
is the AC mains voltage prior to application of the test level.
Note 2: The compliance levels in the ISM frequency range 150 kHz to 2.5 GHz are intended to decrease the likelihood that mobile/por-
table communications equipment could cause interference if it is inadvertently brought into the patient area. However, emission limits,
IEC 60601 test levels, and tests specified in IEC 60601-1-2:2001 do not address electromagnetic compatibility of electrical equipment at
very close distances. Care should always be exercised when using any electrical or RF equipment in the immediate patient area.