14
Miscellaneous
Oil Insulated High Voltage Tanks
The oil-filled tanks in all the PFT SERIES of hipots are field serviceable. The only requirement is that the tank
must be oil filled under vacuum at re-assembly if left out of the oil for longer than 3 hours. The parts to service
the tank are available from HIGH VOLTAGE, INC. at the address noted on the inside front cover of this
manual.
The oil level in the tank should be .5 inches from the lid when the oil temperature is 20
°
C.
Packing the Cables
The output cable will easily fit into the cable storage area next to the control panel. Care should be used to
avoid damaging the output cable jacket when coiling into the compartment. As the final step, the ground and
return leads will coil neatly inside the output cable coil.