Ayrshire

™ User/Service Manual 

 

 

 

 

 

 

 

 

 

 

 

 

 

Herida II 

 

 
 

 

 
 
 
 

NB: Please ensure this document is fully read & understood  

by the Clinician / Carer / Service Engineer prior to use. 

 
 

User & Full Service Manual 

Pressure Relieving / Alternating Air Mattress & Pump 

 

Pump Model: HH/P09  Mattress Model: HH/T04 

 

Summary of Contents for Ayrshire

Page 1: ...ervice Manual Herida II NB Please ensure this document is fully read understood by the Clinician Carer Service Engineer prior to use User Full Service Manual Pressure Relieving Alternating Air Mattres...

Page 2: ...2 P a g e J A N U A R Y 2 0 2 0 Ayrshire User Service Manual...

Page 3: ...e OPERATION 6 Static Mode Alternating Mode Profiled Bed Clinical Considerations Installation Safety Instructions Pump Operation Mattress Pump Characteristics MAINTENANCE 10 Cleaning Maintenance Repair...

Page 4: ...g solutions Keep sharp objects away from the mattress Ensure that the system is clean and dry prior to storage Store the pump and mattress in the protective bags supplied We recommend the use of cot s...

Page 5: ...makes it easy to clean whilst complying with strict infection control procedures The prompt and rapid cell deflation between cycles physically ventilates thus allowing immediate pressure relief betwee...

Page 6: ...chambers deflating for 5 minutes This is one complete cycle Profiled Bed Clinical Considerations If the bed is in the profiled position ensure that the pump pressure is increased accordingly to appro...

Page 7: ...g the mattress laying out flat The air inlet should be located at the feet side of the bed 3 Use the Rapid Inflator if available to inflate the mattress before connecting the air tubes to the pump The...

Page 8: ...Pressure Monitor The light is illuminated when pressure is normal extinguished when pressure is abnormal The light will continuously flash during the alternating cycle change over Mute When pushed th...

Page 9: ...stance to rectify Power Off Alarm The alarm will sound if power is lost to the pump Pump Characteristics Power AC220V AC110V Pressure Range 10 52 mmHg Internal Fuse F0 5A 250V Plug Fuse 3A Frequency 5...

Page 10: ...a Healthcare Ltd and only original manufacturers parts should be used Failure to follow the instructions could result in a void warranty situation Service Interval Instructions This product must be se...

Page 11: ...stem soak test over at least 3 cycles or under weights to replicate patient in situ The tubing and connectors that run down the side of the mattress must be clear of kinks debris or must not be snappe...

Page 12: ...age 9 using a Herida Testing System Calibration training must be conducted applied by a Herida trained technician Other Considerations Original internal fuse is F0 5A 250V When replacing the fuse ensu...

Page 13: ...luding cells tubing and CPR toggle Ensure all cell fasteners are snapped to the mattress base and are not lose or faulty Check that stitching which secures the straps to the base of the mattress is so...

Page 14: ...14 P a g e J A N U A R Y 2 0 2 0 Ayrshire User Service Manual P09 Pump...

Page 15: ...against electric shock Class II electrical appliance Double insulated Dispose of electrical equipment correctly WEEE Do not use broken or damaged electrical plugs or cables as this may lead to electri...

Page 16: ...tors Repair the air tubes or call for replacement Re attach CPR plug 3 Mattress is too hard or too soft The air pressure is not adjusting to required level Adjust the pump pressure according to the us...

Page 17: ...Leeds LS27 0LL Freephone 44 800 193 6030 Email sales heridahealthcare co uk Company Number 9918019 VAT Registration Number 233587691 Disclaimer Please be aware that all products should be disposed of...