Hydrim C61WD Recommissioning Guide                         April 2020 Rev 1.0 

 

Contact our service desk on 

0800 028 4749 

or email

[email protected]

 

Annex A – Changing the air filter and bacteriological filter (if needed) 

When the message “Replace air filter” appears, the HEPA filter needs changing. 

1.  Turn the power OFF 

 

2.  Pull open the ventilation panel below the front 

door of the unit 
 

3.  Make note of the direction of the arrow and 

remove the old filter by pulling it from the centre 
 

4.  Install new air filter – placing the arrow in the 

correct orientation – and close the ventilation 
panel 
 

5.  After replacing the filter, go to the “Reset Drying 

Counter” screen in the Setup menu and reset 
the drying counter to zero  
 

Parts you may need 

Description 

Part Number 

 

Salt  

1132684 

HEPA filter 

1137008 

Bacteria Retentive Air Filter 

1043321 

 

If you are in any doubt 

WHATSOEVER

 as to the condition and/or safe operation of your 

Hydrim C61wd, you should switch OFF any power source and clearly identify the equipment 
as 

DO NOT USE

. Contact Henry Schein Service for support. 