10.2 How to clean the strip holder
Wipe off urine residues from the strip holder with a lint-free cloth after each
measurement. This prevents crustification and drying of urine residues.
The strip holder can be removed from its transport mechanism and should
be cleaned with water and - when necessary - with cleaning agent or
disinfectant. Make sure that the instrument is turned off before removing
the test slide.
After cleaning, the strip holder should be put back onto its transport
mechanism carefully. The rectangular notches of transport mechanism and
stick retainer must be placed on top of each other (Pic. 8.3).
Pic. 8.1 : test slide
(bottom view)
Pic. 8.2 : Notch A
Pic. 8.3: Notch B