13 

 

the ▼ can change the number of the relevant digits. The number can be 

cyclic  changed  from  0~9,  after  this,  you  can  press  the  OK  button  to 

finish  the  demarcated  gas  concentration  setting,  the  interface  of  the 

standard gas concentration as below: 

 

Picture 3.8

Calibration gas concentration interface 

3.2.7

Span calibrations setting: 

It  refers  to  that  the  users  can  adjust  the  instrument  by  using  the 

calibration  gas.  After  entering  this  interface,  press  the  M  button  to 

calibrate,  then  the  SET  PEAK  becomes  flickering.  After  that,  pass 

through the standard gas(control the flow as 200ml/min), when the data 

stable, press the “OK” to save the setting. 

 

 

 

 

 

    Picture 3.9

Span calibration setting 

四、

Calibration methods 

Before  the  gas  detector  leave  our  factory,  we  have  calibrated  it.  If 

the users have the calibration device, they can calibrate it regularly as the 

Summary of Contents for OC-F08

Page 1: ...OC F08 Fixed gas detector Operating Manual Henan Oceanus...

Page 2: ...ny gas detector also should be according to the rules of related department 6 Fixed gas detector should be calibrated 6 months but for some sensor we suggest calibrate every 3 months 7 The sensor is t...

Page 3: ...iring also with the explosion proof design Signal output optional of the 4 20mA RS485 0 5V 0 10V 200 1000Hz relay output Available for the DCS PLC or our control system With high sensitivity quick res...

Page 4: ...ss of usage please keep the air inlet clean or will display lower data Please do not disassemble Please not put gas detector under the high temperature and humidity also the strong static electric pla...

Page 5: ...4 Installation A Site installation B Installation and size...

Page 6: ...erminal GND is 24V negative terminal 3 ALARM connect with 24V audible visual alarm if choose the 200 1000Hz frequency signal it could be frequency output interface 4 K1 is relay output Grade I alarm i...

Page 7: ...d from ground about 0 3 0 6m if light than air should higher the leakage source than 0 5 2m The gas detector should keep right distance from leakage source if too close when the leakage frequency will...

Page 8: ...nd industry standards The DC signal and AC power supply line should not in one shielding cable Gas detector should be connected with the earth through pipeline holder or the cable For the RS485 signal...

Page 9: ...ert into the shell from the bottom cap left hole If need after insert into the accordingly terminal and then fixed the power line and signal line Remark In the explosive dangerous place the user must...

Page 10: ...rmal inspection mode Picture 3 2 Normal inspection interface In the process of inspection when the gas concentration exceed the set level AL1 and AL2 will display meanwhile the red backlight will ligh...

Page 11: ...ain long press the M will back to the normal inspection interface 3 2 1 AL1 setting When the gas concentration data exceed the set level the red backlight will be lit meanwhile the K1 relay will act P...

Page 12: ...address setting Picture 3 5 setting the device address interface Users can set the device address through this interface the range of the address is 1 247 Among the RS485 bus system every device is on...

Page 13: ...press the M button under this interface Picture 3 7 Zero calibration interface The SET ZERO character flickers and the interface display the relevant number at the same time after the number is stable...

Page 14: ...setting It refers to that the users can adjust the instrument by using the calibration gas After entering this interface press the M button to calibrate then the SET PEAK becomes flickering After tha...

Page 15: ...4 1 Zero calibration Pass the 0ppm concentration standard gas through the gas detector usually as the normal air as the zero some gas such as the O2 CO2 take the N2 as the Zero then enter into the Ze...

Page 16: ...lashing after the value is stable don t rise or descend any more press the OK button to save and exit Remote controller Remote controller mainly for the working place with heavy dust or other place in...

Page 17: ...when it working Avoid the dropping from height or severe vibrations Do not regular contact the device with high concentration gas which has higher concentration than the detecting range or the sensor...

Page 18: ...Detecting Range According to the detected gas Resolution According to the detected gas Sampling Diffusion or pump suction circulation selectable Accuracy 3 F S Repeatability 1 Zero drift 1 F S year S...

Page 19: ...M 10 PPM 20 PPM H2 0 1000 PPM 50 PPM 100 PPM CH2O 0 100 PPM 25 PPM 50 PPM O2 0 30 0 vol 18 5 vol 23 5 vol H2S 0 100 PPM 10 PPM 20 PPM HCl 0 30 0 PPM 5 0 PPM 10 0 PPM HCN 0 50 0 PPM 5 0 PPM 10 0 PPM PH...

Page 20: ...LEL 25 LEL 50 LEL CO2 0 5 0 vol 1 0 vol 3 0 vol NDIR infrared sensor 0 5000ppm 800 PPM 1500 PPM CH4 0 5 0 vol 0 100 vol 1 0 vol 3 0 vol 5 vol 10 vol Remark if not listed gas type or measuring range pl...

Reviews: