113
Result Verification
Patient Test
PID: 750629 - 3694
LID: 12345
= 64 mg/dL
1) 63 mg/dL
2) 65 mg/dL
Result Verify
5
1
3
2
4
FIGURE 4-15
1) indicates the result from the first
measurement (1).
2) indicates the result from the second
measurement (2).
x
¯
indicates the Mean value (3) for these two
measurements.
If the Duplicate sampling function is used
and the two results differ more than the
maximum difference set in the settings (see
), the results will automatically
be rejected and instead of the mean value
the following text will be displayed
Difference too large
.
When the confirm button is pressed
is displayed. When pressing the confirm
button (1) in
the Main Menu
will be displayed.
1 Result from first measurement
2 Result from second measurement
3 Mean value
4 Comment input button
5 Confirm button
Duplicate sampling
Warning
1
Difference too large.
Please perform
another
measurement.
FIGURE 4-16
Summary of Contents for 201 DM