Appendix B - i.C
3
Reference Guide
General Settings
>
Notes
• All i.C
3
settings are shown in this guide. Contact Helmer, Inc. for information about settings as they pertain to optional
applications.
•
General settings apply to refrigerator and freezer models.
Setting
Description
Default Value
Range
Temperature Setpoints
Set unit operation temperature
22°C
20 °C - 35 °C
Alarm Settings
Access alarm setpoint settings
Refer to Alarm Settings table
Password Protection
Protect i.C
3
Settings from unauthorized changes
On
Change Password
Change password used to access i.C
3
Settings
1234
Sounds
Turn sound on or off
(does not include alarms)
On
Alarm Volume
Audible alarm volume
9
1 - 9
Alarm Tone
Audible alarm tone
3
1 - 4
Language
Language displayed on the i.C
3
screen
English
Date Format
Date display format
MM/DD/YYYY
Day
Calendar day
Set at the factory
Month
Calendar month
Set at the factory
Year
Calendar year
Set at the factory
Time Format
Time display format
12-hour
12-hour or 24-
hour
Minute
Minute value
Set at the factory
Hour
Hour value
Set at the factory
AM/PM
AM or PM
Set at the factory
Temperature Units
Temperature units in °C or °F
°C
°C or °F
Temperature Graph Screensaver Display temperature graph as Home screen
On
Alternate Screensaver
(optional)
Display alternate screensaver as Home screen
Off
Unit ID
Unique identification number for the i.C
3
system
Unit serial number
Max 10
characters
Min/Max
Minimum and Maximum Primary Monitor Probe readings
Off
Access Control as Home Page
Display Access Control keypad on the Home screen
On
(if installed)
Brightness
i.C
3
display screen brightness
3
(brightest)
1 - 3
Sensor Calibration
Set probe temperature values to match measured
temperature values
Refer to product Service Manual
Primary Monitor Probe Offset
Offset value used to adjust temperature reading of
the primary monitor temperature probe.
0°C
-10 °C - +10 °C
-18 °F - +18 °F
Control Probe Offset
Offset value used to adjust temperature reading of
the control probe
0°C
-10 °C - +10 °C
-18 °F - +18 °F
Ambient Probe Offset
Offset value used to adjust temperature reading of
the ambient probe.
0°C
USB Port
Turn USB port on or off
On
Ethernet Port
Turn Ethernet Port on or off
Off
Restore Factory Settings
Change all settings to factory default values
Refer to product Service Manual
Helmer Scientific
i.C
3
User Guide for i.Series Incubators
360406/A