360179-A/B
24
Installation & Configuration
8.4.2
Change Sensor Calibration Offset Values
NOTE
►
Contact Haemonetics
Corporation BloodTrack
Customer Support for instructions on
changing sensor calibration offset values.
8 .5
Factory Default Settings
Factory settings may be simultaneously returned to factory default values.
NOTE
►
The Factory Default Settings may not be the same as the settings that were factory-
calibrated before the Compartmental Access Refrigerator was shipped.
►
If Factory Default Settings are restored, all calibration values and User PINs will be
lost.
General Settings
Restored Value
Home Screen Application Icons
i.C³ APPS, Settings, Temperature Graph, Automatic
Alarm Test, Information Logs
Temperature Setpoint
4.0 ºC
Password (for Settings screen)
1234
Password Protection (for Settings screen)
On
Sounds
On
Alarm Volume
9
Alarm Tone
2
Unit ID
No default value
Date Format
MM/DD/YY
Day
Not affected (maintained in real-time clock)
Month
Year
Time Format
12-hour
Minute
Not affected (maintained in real-time clock)
Hour
AM/PM
Language
English
Temperature Units
ºC
Temperature Graph Screensaver
Off
Alternate Screensaver
On
Min/Max Temperature Display
On
Air Probe Temperature Display
Off
Air Probe Temperature Alarm
Off
Display Brightness
High (3 symbols)
Temperature Calibration Settings
Values entered at the factory
Access Control Touchpad
On
Alarm Settings
Restored Value
Upper Probe High Temperature (Setpoint)
5.5 ºC
Upper Probe High Temperature (Time Delay)
0 minutes