Helios PD2-6400 High Voltage & High Current Meter
Instruction Manual
38
Advanced Features Menu
To simplify the setup process, functions not needed
for most applications are located in the
Advanced
Features
menu.
Press and hold the Menu button for three seconds to
access the advanced features of the meter.
Advanced Features Menu &
Display Messages
The following table shows the functions and
messages of the
Advanced Features
menu in the
order they appear in the menu.
Advanced Features Menu & Display Messages
Display
Parameter
Action/Setting
filtEr
Filter
Set noise filter value
Ch-A
Channel A
Set filter value for channel
A
Ch-V
Channel V
Set filter value for channel
V
bypASs
Bypass
Set filter bypass value
Ch-A
Channel A
Set filter bypass value for
channel A
Ch-V
Channel V
Set filter bypass value for
channel V
Round
Round
Set the rounding value for
display variables
seriAl
Serial
Set serial communication
parameters
SlaVId
Slave ID
Set slave ID or meter
address
Baud
Baud rate
Select baud rate
Tr dLY
Transmit delay
Set transmit delay for serial
communication
Parity
Parity
Select parity
Even, Odd, or None with 1
or 2 stop bits
t-byt
Time byte
Set byte-to-byte timeout
SelEct
Select
Enter the Select menu
(function, cutoff, out)
Functn
Input signal
conditioning
Linear
Ch-A
Channel A
Select menu for channel A
Ch-V
Channel V
Select menu for channel V
Linear
Linear
Set meter for linear function
and select number of
linearization points
No pts
Number of
points
Set the number of
linearization points (default:
2)
nmath
Math
Enter the setup menu for
channel P math function
nmulti
Multiplication
Channel P = ((A*V)+C)*F
Const
Constant
Constant used in channel P
math
Advanced Features Menu & Display Messages
Display
Parameter
Action/Setting
adder
Adder
Addition constant used in
channel P math calculation
(C)
factor
Factor
Multiplication constant used
in channel P math
calculation (F)
CutofF
Cutoff
Set low-value cutoff
Ch-A
Channel A
Set low-value cutoff for
Channel A
Ch-V
Channel V
Set low-value cutoff for
Channel V
AoutPr
Analog output
programming
Program analog output
parameters
AOut 1*
Analog output 1
Program analog output 1
parameters
Source
Source
Select source for the 4-20
mA output
break
Current input
break
Set relay condition if current
input break detected
O-rang
Overrange
Program mA output for
display overrange
u-rang
Underrange
Program mA output for
display underrange
nmAx
Maximum
Program maximum mA
output allowed
nmin
Minimum
Program minimum mA
output allowed
Calib
Calibrate
Calibrate 4-20 mA output
(internal reference source
used for scaling the output)
4 nmA
4 mA output
Enter mA output value read
by milliamp meter with at
least 0.001 mA resolution
20 nmA
20 mA output
Enter mA output value read
by milliamp meter with at
least 0.001 mA resolution
User
User I/O
Assign function keys and
digital I/O
F1*
F1* function key
Assign F1 function key
(*F1/F2/F3)
F4
F4 function
Assign F4 function (digital
input)
dI I
Digital input 1
Assign digital input 1
– 4
dO 1
Digital output 1
Assign digital output 1
– 4
ICal
Internal
calibration
Enter internal calibration
(used for recalibrating the
meter with a calibrated
signal source)
Ch-A
Channel A
Perform calibration on
channel A
C CAL
Current
calibration
Calibrate current input
(internal reference source
used for scaling the input)
C lo
Current low
Calibrate low current input
(e.g. 0 A)