EX-Supplementary Operating Manual KDSE 

January 2021 

 

 

Electronic transmitter

 

KDSE 

 

EX-Supplementary Operating Manual for Hazardous areas  
Safety Instructions 

 

 

 

 

 

 

 

 

 

 

 

 

This operating manual is a supplement to the manuals for the  

Flow-meters KDS und BGK. 

 

Please read the instructions carefully and store them in a safe place.

 

Summary of Contents for KDSE

Page 1: ...nuary 2021 Electronic transmitter KDSE EX Supplementary Operating Manual for Hazardous areas Safety Instructions This operating manual is a supplement to the manuals for the Flow meters KDS und BGK Pl...

Page 2: ...aintenance 4 1 2 Hazard warnings 5 1 3 Proper use of the device 5 2 Identification 6 3 General information about explosion protection 7 4 Applications 8 5 Operational mode and system design 8 6 Electr...

Page 3: ...ruc tion February 2021 These instructions are supplementary operating manual for non explosion proof meters If you do not have a copy of the latter instructions please request one from Heinrichs Messt...

Page 4: ...y maintenance or repair which is safety relevant in terms of explosion protection is to be carried out by the manufacturer an authorized Heinrichs Messtechnik GmbH service center or under the supervis...

Page 5: ...l material damage or even death Always comply to these warnings and proceed with caution Caution means that failure to take the prescribed precaution could result in material damage or de struction of...

Page 6: ...hs Messtechnik GmbH Robert Perthel Strasse 9 D 50739 Cologne Germany Phone 49 221 49708 0 Fax 49 221 49708 178 Internet www heinrichs eu Email info heinrichs eu Product type Electronic transmitter for...

Page 7: ...o occur 3G 2 3D 22 Equipment in this category is intended for use in areas in which explosive atmospheres caused by gases vapours mists or air dust mixtures are unlikely to occur or if they do occur a...

Page 8: ...ion of a magnet in the variable area flow meters series KDS and BGK Either the position of the float in which there is a permanent magnet or the angular position of a ring magnet which is located on a...

Page 9: ...as described below 6 1 Electrical and thermal parameters Voltage Ui DC 30 V Current Ii 150 mA Effective internal inductance Li 0 24 mH Effective internal capacitance Ci 15 nF Power Pi in acc with the...

Page 10: ...g tube 7 3 Ground connection In variable area flow meters it is possible under operating conditions for charge separation to occur in the measuring tube due to the transport of non conductive fluids a...

Page 11: ...age 11 of 14 8 Marking Warning Only devices with ex marking may be operated within the explosive atmospheres range The electronic module is marked as follows When the module is assembled in a variable...

Page 12: ...Heinrichs Messtechnik GmbH Page 12 of 14 Ex Supplement operating manual KDSE 9 Declaration of Conformity...

Page 13: ...Heinrichs Messtechnik GmbH Ex Supplement operating manual KDSE Page 13 of 14...

Page 14: ...Heinrichs Messtechnik GmbH Page 14 of 14 Ex Supplement operating manual KDSE...

Reviews: