Warranty

Instead of the usual 2-year guarantee, we guarantee this device for 5 years from the date of delivery 
from our factory (this excludes consumables such as bulbs, tips and batteries).
We guarantee the proper function of the unit provided it is used as intended by the manufacturer and 
according to the instructions for use. Any faults or defects which occur during the guarantee period 
will be repaired free of charge provided they are caused by faults in material, design or workmanship. 
In the case of a complaint of a defect in the product during the guarantee period, the buyer must prove 
that any such defect was present at the time the product was delivered. The usual legal guarantee 
and warranty do not apply to incorrect use, the use of non-original HEINE parts or accessories. 
(Especially bulbs, as these are designed with emphasis on the following criteria: Color temperature, life 
expectancy, safety, optical quality and performance.) Also excluded are repairs or modifications made 
by persons not authorised by HEINE or cases where the customer does not follow the instructions 
supplied with the product. Any modification of a HEINE product with parts or additional parts which do 
not conform to the original HEINE specification will invalidate the warranty for the correct function of 
the product and further invalidate any warranty claims which result from such a change or modification. 
Further claims, in particular claims for damage not directly related to the HEINE product, are excluded.

For 

repairs

, please contact your supplier, who will return the goods to us or our authorized agent.

Bought from

date

Instructions for use 

HEINE i-View

®

Before using the HEINE i-View

®

, read these instructions carefully and keep them in a safe

place for future reference.

Safety information

Range of applications: 

The HEINE i-View

®

 is designed for use exclusively with HEINE instruments

and light sources.

General Safety:

 Use only HEINE original replacement parts and accessories.

Safety in use:

 To avoid possible damage to the eyes, avoid looking directly at the sun or a bright light!

The adaptors in illustrations 1+2 must be assembled as shown. HEINE declines responsibility for any
assembly defects caused by the optician.
Do not unscrew the set screws. They are factory-set for the appropriate working distance.

Description:

 The HEINE i-View

®

 and the HEINE S-Guard

®

 are already assembled by HEINE comple-

tely. Attach the HEINE i-View

®

 to the headlight using the adaptors which are already assembled by 

HEINE, too. The ease of adjustment of the joints shown in illustrations 1+2 can be adjusted with the 
screwdriver supplied.

Adjustment and Operation

Adjusting the nose bridge (A):

 This can be bent into position to suit the anatomy of your nose.

By bending the bridge supports you can vary the distance between the lenses and the eyes.
This avoids contact between your eyelashes or cheekbones and the frame or corrective lenses.

Changing the nose pad:

 The nose pad is attached to the S-Guard by a screw. To change the 

nose pad, unscrew the screw and remove the old nose pad.

Fitting optional corrective lenses:

 Please consult an optician before having corrective lenses fitted.

Please show the optician the attached “Instructions for specifying corrective lenses”. To fit the
corrective lenses (B), insert one side of the snap-in frame (2) into the groove in the S-Frame. Next,
insert the other side of the snap-in frame in the second groove.
To remove the snap-in frame with corrective lenses, reverse the above procedure, avoiding damage
to the lenses.

Changing protective lenses:

 Hold the S-Frame as in (D) and bend the lens in the direction shown

and pull the lens with an audible click out of the frame. To replace, follow the procedure shown
in (E) and (F) until the lens clicks into place.
If the standard protective lenses are too close to your cheekbones, you can either fit the small
protective lenses or adjust the nose bridge (A).

Swivelling the loupe:

 The HEINE i-View

®

 can be swivelled upwards with one hand. Use the

sterilizable swivel levers to do this (see accessories). This movement makes it possible to observe
the illuminated field under investigation directly without touching the setup of the headlight.

Instructions for specifying corrective lenses:

When specifying and fitting corrective lenses, be sure to take into account the following:
•  The magnifying optics should be attached to the frame to suit your usual working posture.
•  Prismatic errors should be corrected.
•  The corrective lenses are mounted at an angle to the optical axis.
Check the correction by means of an eye test with the magnifying optics and the corrective
lenses fitted.
The total spherical correction is usually max. +/- 5 dpt. Ask your optician or ophthalmologist!
The corrective lenses should be placed as closely as possible to the eyepieces of the magnifying
optics.

Cleaning and Maintenance

General: The plastic protective lenses of the S-Guard splash guard have an anti-reflex coating. Clean 
them carefully to avoid mechanical or chemical damage. 
Manual Cleaning: Clean the S-Guard and the i-View optics mount with the enclosed HEINE microfibre 
cloth. We recommend the HEINE cleaning spray C-000.32.542

 

as a cleaning agent.

Warning: 

 Immersion disinfection, machine cleaning/disinfection, ultrasound cleaning and sterilisation 

are not permitted. 

Specification

Protective lenses: 

Plastic, anti-reflex coated

i-View and S-Guard: 

Nickel-free