Commissioning | Configuring the axes
Reading deviations from the calibration standard
The manufacturer of the calibration standard usually provides you with
specifications about the deviations.
Prerequisites:
An ACF file contains the nominal values; this file must match the import
scheme of the product
"Creating an ACF import file", Page 99
Nonlinear error compensation (NLEC)
is deactivated
Tap
Settings
in the main menu
Open in succession:
Axes
General settings
Error compensation
Nonlinear error compensation (NLEC)
Tap
Read deviations of calibration standard
Navigate to the desired folder
Tap the desired file (.ACF)
Tap
Select
The nominal values are imported from the file
Creating an ACF import file
To import the calibration data into the product, you must acquire these in an ACF
file.
Open a new file in the text editor of your computer
Save the file under a unique name and the *.acf file extension
Enter the values, separated by tab stops as shown in the schema below
The following specifications apply to import files:
File names must not contain diacritics or special characters
Use a point as decimal character
7
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020
99
Summary of Contents for GAGE-CHEK 2000 Series
Page 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 08 2020 ...
Page 12: ......
Page 13: ...1 Fundamentals ...
Page 22: ......
Page 23: ...2 Safety ...
Page 28: ......
Page 29: ...3 Transport and storage ...
Page 34: ......
Page 35: ...4 Mounting ...
Page 41: ...5 Installation ...
Page 51: ...6 Basic operation ...
Page 74: ......
Page 75: ...7 Commissioning ...
Page 119: ...8 Setup ...
Page 154: ......
Page 155: ...9 Measuring ...
Page 169: ...10 File management ...
Page 176: ......
Page 177: ...11 Settings ...
Page 210: ......
Page 211: ...12 Servicing and maintenance ...
Page 220: ......
Page 221: ...13 What to do if ...
Page 225: ...14 Removal and disposal ...
Page 227: ...15 Specifications ...