Commissioning | Configuring the axes
Activating the nonlinear error compensation
Tap
Settings
in the main menu
Open in succession:
Axes
General settings
Error compensation
Nonlinear error compensation (NLEC)
Activate
Compensation
with the
ON/OFF
slide switch
The error compensation is applied from the next
measurement
Configuring
Squareness error compensation (SEC)
Squareness error compensation (SEC)
enables angular errors to be compensated
during measuring point acquisition. The compensation factor is determined by
comparing the deviation of the actual measurement result from the nominal angle
of the spatial axes. The compensation factor is applied to the entire measuring
range.
Tap
Settings
in the main menu
Open in the sequence
Axes
General settings
Error compensation
Squareness error compensation (SEC)
The measured values (M) and nominal values (S) of the three
spatial axes are displayed
Enter the measured values of the reference standard
(= nominal values)
Activate
Compensation
with the
ON/OFF
slide switch
Squareness error compensation is applied from the next
measurement
"Squareness error compensation (SEC)", Page 195
7
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020
111
Summary of Contents for GAGE-CHEK 2000 Series
Page 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 08 2020 ...
Page 12: ......
Page 13: ...1 Fundamentals ...
Page 22: ......
Page 23: ...2 Safety ...
Page 28: ......
Page 29: ...3 Transport and storage ...
Page 34: ......
Page 35: ...4 Mounting ...
Page 41: ...5 Installation ...
Page 51: ...6 Basic operation ...
Page 74: ......
Page 75: ...7 Commissioning ...
Page 119: ...8 Setup ...
Page 154: ......
Page 155: ...9 Measuring ...
Page 169: ...10 File management ...
Page 176: ......
Page 177: ...11 Settings ...
Page 210: ......
Page 211: ...12 Servicing and maintenance ...
Page 220: ......
Page 221: ...13 What to do if ...
Page 225: ...14 Removal and disposal ...
Page 227: ...15 Specifications ...