8
Getting Started
Warning!
Use original battery charger only!
The use of a different charger could damage the battery or
product and would void the warranty. Confirm that the voltage of
the adapter matches the AC voltage of the electrical outlet.
Before starting to charge, confirm that the voltage and the plug of the
AC adapter is identical to the main power supply at the outlet. See
below if the plug type is not identical and needs to be changed.
1. The temperature range for recharging is 50 to 95°F (10 to 35°C).
2. Plug the Micro USB connector into the base of the instrument. Or
insert the Micro USB connector into the stand and place the
instrument on the stand.
The Micro USB connector can only be inserted one way into the unit.
Forcing the plug into the housing will damage both Micro USB and
socket, thus voiding the warranty.
3. Insert the plug of the AC adapter into a suitable wall outlet.
4. Allow to charge up to 6 hours from empty to fully charged. The green
light blinks when charging.
5. Once the instrument is fully charged, the adapter automatically
switches to a trickle charge operation and the red indicator will turn
off.
Remove the contents from the package, examining them carefully for
breakage, defects, or shortages. Save the packaging.
Charging the Battery
For transport safety reasons, the instrument is shipped with minimal
battery charge. The battery must be charged before initial use.
Notes
— When fully charged, the instument can be operated for approximately 20 hours.
— If using the Pipette Controller daily, the product can be charged over night or
any convenient time. A residual charge of approximately 10% is available when
the red LED has lit.