Hearing Technologies Pro-Power Vacuum-Aspirator Setup Download Page 2

 
Gently place the tip of the suction tip in the receiver (earmold) or microphone port or over the microphone 
screen.  
 

If there is a wax-spring in the receiver tube do not force the suction tip into the receiver tube.  
 

If needed attach the canal of the hearing aid to the open suction tube and let it run for 5 minutes.  Try to loosen 
the debris again with a brush and wax-pick.  
 
Repeat using the suction tips if needed. 

 

Maintenance 

 

Clean the Suction Canister, suction tube, suction tips, and cleaning tools with a germicide or 
antibacterial cleaner on a regular basis.  

 

Follow safety protocols for the disposal of used germicide. 

 

Read the instruction of your cleaning solution for safety issues and to confirm that it can bee used on 
plastic and rubber. 

 
The filter that is connected to the stainless steel connector should be changed when the suction power is low 
or is wet from a liquid bypassing the suction canister.   
 

Trouble Shooting 

 

 
 
 
Problem 

 
Pump will not run when switch 
is pushed ON. 
 
 
 
Low Suction 
 
 
 
 
No Suction 
 
 

 
 
 

Probable Cause 

 
Power source 
Fuse is burned out 
 
 
 
Loose tubing connection 
Canister lid loose 
Filter clogged 
 
 
Disconnected tubing 
Filter clogged 
 

 
 

Remedy 

 
Check power outlet 
Check and replace fuse that is 
located under the base of the 
unit. 
 
Check all connections 
Tighten lid 
Replace filter 
 
 
Check all connections 
Replace filter

Contact Hearing Technologies or your accessory distributor for additional assistance. 

Limited Warranty 

Do not use as an Aspirator unless you have been trained to remove cerumen and have 
the necessary attachments that are available from your equipment distributor.  

 

The Pro-Power

 Vacuum- Aspirator vacuum pump is warranted against defect in materials or workmanship 

for 12 months from the date of purchase from Hearing Technologies International, Inc or an authorized 
distributor. Tubing, filters, suction tips, fuses, cleaning tools and shipping charges for repair are not covered by 
this warranty. Do not immerse in water, any misuse of this product voids the warranty.  
 
Should service be required contact Hearing Technologies by email 

[email protected]

 or fax 727-526-0382 for 

instructions to return your unit for repair. All units must be shipped prepaid to Hearing Technologies. 
 

 Hearing Technologies International, Inc 13799 Park Blvd #253 Seminole, Florida 33776 

(800) 553-6003 Fax (727) 526-0382 www.hearingtech.com 