8) Position the extension tube to a
comfortable position by manually
extending or retracting the tube .
9) Slide the variable steam control down for
small-sized particles of steam. Smaller
particles are ideal when inhaling steam
through the nose . Slide the control up
and large-sized particles will result.
Larger particles are ideal when inhaling
through the mouth .
NOTE:
Extending the tube also lowers the
temperature of the steam .
10) Check that the POWER switch is in the
‘off’ position . Plug the power cord into a
main outlet .
11) Turn the POWER switch to the ‘ON’ position.
12) Start your treatment .
13) The Steam Inhaler will begin to output
steam in approximately 60 seconds .
Once the level of steam becomes weak
(approximately 6-9 minutues), turn the
POWER switch to the ‘OFF’ position and
unplug the power cord from the outlet .
IMPORTANT!
The Inhaler will automatically shut off if it
becomes overheated . If/when this happens,
immediately:
1) Turn the POWER switch to the ‘OFF’
position .
2) Unplug the power cord from the outlet .
3) Allow the unit to cool for approximately
30 minutes .
Before restarting the unit, make sure that the air
vents on the extension tube are not obstructed .
After Every Use
1) Unplug the unit from the power source .
2) Allow the unit to completely cool
(approximately 10 minutes) .
3) Carefully remove the aromatherapy/drain
tank and pour out any remaining water .
4) Follow the cleaning procedures provided in
this guidebook .
NOTE:
Occasionally the nozzle that directs the
steam into the extension tube may become
obstructed . If the Steam Inhaler begins to
produce a significanty smaller amount of steam
or no steam at all, the nozzle needs to be
cleaned. To clean the nozzle, remove the cover
from the Steam Inhaler. Firmly grasp the nozzle
and twist clockwise to remove,
Fig. 7.
Insert a
fine gauge needle or the pin which is stored
under the bottom of the Steam Inhaler into the
nozzle to clear the airway,
Fig. 7
. To reattach the
nozzle, align the nozzle brackets with the
grooves on the nozzle base. Turn
counterclockwise until the nozzle locks into
place .
Troubleshooting
If any abnormality occurs during use, please
check and correct the following:
Condition Correction
Unit does not
Check the AC connection
operate when
to the outlet .
POWER switch
is turned .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
No steam
Check that there is water
or low rate
in the water heater tank .
of steam
Check that there is
saline or aromatherapy
water in the
aromatherapy/drain tank
(if applicable) .
Make sure that suction
tube and other components
are properly attached .
Make sure that the steam
nozzle is clean. Remove
nozzle and clear the airway
with the pin or fine gauge
needle .