Healthmark TOSI WT101 Instructions For Use Download Page 2

(

for additional test procedures with the TOSI

®

, please visit: 

http://www.hmark.com/tosi.php

1.

 

Open the TOSI

®  

package by tearing across the precut notch at the top of  the TOSI

®

 

package; remove the TOSI

® 

coupon from the package.  

2.

 

Secure one (1) TOSI

® 

per level in the center of an empty tray as depicted in the 

diagram above. 

3.

 

Place one on each shelf.  In multiple shelf units follow diagram above and examples: 
e.g. 3 Shelf Unit, 3 TOSI

®

; 4 Shelf, 4 TOSI

® 

or 1 Level Tunnel Washer, 1 TOSI

®

.) 

4.

 

Process using your surgical instrument wash cycle. 

5.

 

Examine the TOSI

® 

for visual cleanliness.  Compare the test to the 

interpretation chart

WT104, scale (0-5).  

6.

 

Record results on 

log sheet

 on www.hmark.com.  

 

Sonic & Blood Soil Test 

1.

 

Make sure that the ultrasonic cleaner has been degassed prior to running the test 
and has the correct amount of cleaning solution in the tank/bath. 

2.

 

The number of SonoChecks placed in the tank will be dependent on the volume of 
the tanks (see routine testing). 

3.

 

Place the correct amount of SonoChecks in the appropriate tray for testing 
according to tank size. 

4.

 

Secure one (1) TOSI

®

 to the middle of the tray. 

5.

 

TOSI

®

 and SonoCheck

TM

 can be placed in the same tray. 

6.

 

Run the Sonic through its normal cycle (record the cycle). 

7.

 

Record both the SonoCheck

TM

 and TOSI

®

 results at the end of the cycle. 

Interpretation of Results 

If less than optimal results were obtained (1-5), make adjustments to the equipment 
utilizing the 

chart (WT104)

 as a guide. Typical failures can be caused by: 

 

clogged spinner arms 

 

worn spray arm bushings 

 

insufficient detergent due to empty containers 

 

detergent pump failure or clogged delivery tubing 

 

poorly functioning water pump 

 

incorrect water temperature 

 

incorrect cycle settings 

Contraindications of Test Results 

 

Tiny Red Spot on TOSI

®

 Plate

: Very rarely, but nonetheless possible, is a slight 

imperfection in the stainless-steel plate which leads to oxidation of the metal. The 
result is a little red speck which could be confused with the hemoglobin soil on the 
TOSI

®

. The easiest way to double check for such a thing is to directly employ 

mechanical action (with a gloved hand, preferably with the aid of an instrument 
cleaning brush) under water. If the speck remains, then it is definitely not the TOSI

®

 

blood soil that remains.  

 

Ghosting on the TOSI

®

 Plate

: A whitish stain is observed on the TOSI

®

 plate, which 

can be confused with fibrin protein remaining on the TOSI

®

. This usually happens at a 

facility that has hard water. If allowed to dry and the TOSI

®

 is read at that point, hard 

water staining may be observed on the TOSI

®

 plate. The simplest method is to gently 

submerge the TOSI

®

 plate in a bath of water. If the stain “disappears” when wetted, 

this indicates a non-test soil residue (likely hard water minerals or detergent) and not 
the blood soil. 

 

Ultrasonic Cycle: Residual Soil Re-deposited on TOSI

®

 Plate: 

In ultrasonic 

washers that do not have a distinct rinsing stage within the same tank (draining of 
cleaning solution and filling with a fresh bath of water), it is possible that a small 
amount of TOSI

®

 soil can become trapped underneath the clear plastic holder and that 

soil is then redeposited upon the coupon. To determine if that is the case, simply give a 
gentle agitation of the TOSI

®

 within a bath. Any soil that was detached during the 

cycle should be rinsed away. If soil still remains attached to the surface of the TOSI

®

 

after this rinsing procedure, then this indicates sub-optimal cleaning. 

Documentation

 

Record all changes and adjustments to the washer according to the results found from the 
TOSI

® 

test relative to the chart scale. Utilize the supplied log sheet. 

Special Warnings and Cautions 

 

Disposal 

Since the TOSI

® 

is run in the mechanical cleaning equipment, there is a chance for 

contamination. Therefore, it is recommended to dispose of the used TOSI

®

 in a biohazard 

container in compliance with facility protocols. 

 