TROUBLE SHOOTING GUIDE 

 

If the SonoCheck

TM 

does not change color or if the time required to generate the color change takes longer than normal, 

please check the following guide: 

 
 

 

 

Problem

 

 

Reason

 

 

Corrective action

 

De-gassing

 

 

Dissolved gasses will absorb ultrasonic energy

 

 

De-gas solution according to equipment man-
ual

 

Water level

 

 

Ultrasonic energy may reflect off of the surface of the 
solution and change energy distribution

 

 

Check equipment manual for correct water 
level

 

Operating cycle time

 

 

Time varies with the amount of ultrasonic energy 
available

 

 

Longer operating cycles generally provide bet-
ter results

 

Instrument load

 

 

Heavy instrument loading and certain materials can 
absorb ultrasonic energy

 

 

Look for weak points using the periodic func-
tional test and check for ultrasonic absorbent 
material like silicone or plastics

 

Transducer failure

 

 

Transducer efficiency may decrease with age.  Indi-
vidual transducers may fail while others in the equip-
ment continue to function

 

 

Perform periodic functional test, placing Sono-
check monitors in each transducer location 
(see equipment manual)

 

Low energy

 

 

Transducer inefficiency or the ultrasonic basket may 
absorb too much energy

 

 

Check performance without basket in place.  
Compare performance against another ultra-
sonic cleaner if available.  Call for service

 