  LOG SHEET 

 

 

ULTRASONIC CLEANER TEST 

 

Machine Number: ________________     Normal Cycle Time:___________ Facility Name:____________________  Date Started:__________ 

Quick Steps for testing Sonic Cavita-
tion using the SonoCheck

TM

 

 

1. 

Prepare a fresh tank of the cleaning 

solution. 
 
2.    Turn the unit on and de-gas accord-
ing to the manufactures instructions 
 
3.  When de-gassing is done, place the 
SonoCheck

TM 

in the center of an empty 

tray / basket and place in the sonic 
cleaner. 
 
4.  Turn the unit on and set to normal 
cycle time. 
 
5.  Start cycle 
 
6.  Observe the SonoCheck

TM

 for a 

color change from blue to yellow (color 
change could happen as quickly as 20 
seconds). 
 
7.     Record the time it takes for the 
SonoCheck

TM

 to change color from blue 

to yellow on the log sheet 
In the case of no color change in the 
normal cycle time refer to trouble shoot-
ing guide.  

Interpretation 

FAIL 

Color remains blue 

PASS 

Color changed to yellow 

Date of Test 

Ex.1/1/06 

 

 

 

 

 

Detergent Used in 

Solution 

(Product Name) 

 

 

 

 

 

Degassing of   

Solution 

Y   or   N 

 

 

 

 

 

Time required to 

have a color 

change to yellow 

 

 

 

 

 

Comments 

 

 

 

 

 

Person Performing 

the Test Name 

 

 

 

 

 

If the SonoCheck

TM 

does not change color or if the time required to generate the color change takes longer than normal, please check the trouble shooting 

guide: 