3

 

TROUBLESHOOTING (CONTINUED) 

 

ERROR ID 

DEFINITION 

CHECK/REPLACE 

Err 1 

Mathematical mistake; division by zero.  

Cannot calculate calibration factor. 

The calibration is performed with no calibration weight 

Err 2 

Weight loaded too low.   
Cannot calibrate. 

While in calibration procedure add more weight in load 

mode 

Err 3 

Unstable weight.  Cannot calibrate 

Check the load cells mechanical surroundings and see 

nothing touches them and the cables are well connected 

Err 5 

Calibration disabled 

Toggle the calibration switch to enable 

Err 6 

High saturation state (high A/D) 

The load cell is not connected correctly.  Check the 

connection, cables, and mechanical connection.  If  the 
problem persists, replace the set of load cells 

Err 7 

Low saturation state (low A/D) 

See Err 6 

 

CALIBRATION OF THE 753KL 

 

This scale has been factory calibrated and does not require calibration prior to use, but 

requires official inspection prior to use as a legal for trade scale. 

 

On the back of the display module, to the left of the DC IN power jack, is a small access cover plate.  

The cover plate is held on by a hex sealing screw.  Removing the cover plate exposes the Calibration 

Enable / Disable or push off / push on switch.  The scale is shipped with the switch in the calibration 

disable position (push on). 

 

The calibration procedure may be performed using a known weight in either pounds or kilograms. The 

default is in pounds, but calibration can be done according to the unit of measure used upon entering into 

the calibration path.  The following instructions assume the scale is set to measure lbs. If you prefer to 

calibrate in kilograms, please call Health o meter® Professional Technical Support at 800-638-3722.     

Note

:  Calibration weight must be no less than 60 lb (27kg) and no more than 300 lb (150 kg).  

 

ACTION 

DISPLAY SHOWS 

Enable scale calibration by toggling the Calibration push switch off. 

 

While display module is turned on and “START’ is  displayed on the LCD (for 

3 seconds); Press ZERO-kg/lb-kg/lb-ZERO 

“CAL” 

Press ENTER 

 “Load/300.0” 

Press ENTER to calibrate with 300lb.  Press BMI to change the weight and 

then the arrows to change the flashing digit, BMI to move to the next digit, 

until you reach the required weight (no less than 60lb /27kg and no more 

than 300lb / 150kg) 

“CLEAR” 

Clear the platform and press ENTER 

 “-----------” 

Wait without touching or shaking the scale until the zero calibration process 

is finished 

“Put XXX.X” 

Load the platform with the required weight and press ENTER 

“-----------” 

Press ENTER 

“DONE”/“CLEAR” 

Remove weight from the platform 

“START” 

Wait until the scale resumes normal operation 

“0.0” 

Disable scale calibration by toggling the Calibration push switch on. 

 

Replace the access cover plate. 

 

Visit http://www.balances.com for your next Healthometer Scale

http://scaleman.com/health-o-meter.html offers the complete line of Healthometer Medical Scales

Summary of Contents for 753KL

Page 1: ...REMOTO BALANCE ELECTRONIQUE PROFESSIONNEL AVEC AFFICHAGE DISTANCE USER INSTRUCTIONS INSTRUCCI NES DE USO MODE D EMPLOI NTEP CC 09 029 Rev 20100618 Visit http www balances com for your next Healthomete...

Page 2: ...PA OL 11 FRAN AIS 18 CAUTION AND WARNING 2 SPECIFICATIONS FOR THIS SCALE 2 QUICK START INSTRUCTIONS 3 ASSEMBLY INSTRUCTIONS 4 SET UP INFORMATION BATTERIES 4 MAINTENANCE 5 TROUBLESHOOTING 5 CALIBRATION...

Page 3: ...s Certified by NTEP to applicable requirements of NIST Handbook 44 for commercial or professional use Your precision scale is designed to provide accurate reliable and repeatable measurements can tran...

Page 4: ...anding still while on the scale Obtaining a Body Mass Index BMI Reading 1 Obtain a weight as explained above 2 Press the BMI button on the keypad the scale will prompt you to enter the person s height...

Page 5: ...r to the troubleshooting instructions If the problem is not corrected call Health o meter Professional technical support at 1 800 638 3722 Figure 1 Replacing Batteries Using the Keypad 4 KEY FUNCTION...

Page 6: ...aners TROUBLESHOOTING Refer to the following instructions to check and correct any failure before contacting service personnel SYMPTOM POSSIBLE CAUSE CORRECTIVE ACTION Scale does not turn on 1 Dead Ba...

Page 7: ...performed using a known weight in either pounds or kilograms The default is in pounds but calibration can be done according to the unit of measure used upon entering into the calibration path The foll...

Page 8: ...3 EXPLODED VIEW OF 753KL SCALE 7 Visit http www balances com for your next Healthometer Scale http scaleman com health o meter html offers the complete line of Healthometer Medical Scales...

Page 9: ...4611 00 MAT 1 18 GRADIENTER LEVEL FOR L F T 1 19 T16 087121 00 STANDING PLATFORM 1 20 FAE 250LB 2 LOAD CELL 2 21 T16 087211 00 RUBBER BUMPER D25X25 4 22 BASE OF 753KL 1 23 FLAT HEAD SCREW SOCKET BUTTO...

Page 10: ...3 Visit http www balances com for your next Healthometer Scale http scaleman com health o meter html offers the complete line of Healthometer Medical Scales...

Reviews: