background image

20 

OPERATING INSTRUCTIONS (CONT.)

 

Parent/Child Function 

The Parent/Child function on Health o meter

®

 Professional scales provides the 

convenience of weighing a small child or infant who is being held by a parent. Follow 
these simple steps to subtract the parent’s weight from the total so only the child’s 
weight is displayed.  

 

1. 

Wait until “0.0” appears on the display. 

 

 

2.  Ask the parent to step onto the scale without the child. 

The display will show a value for the parent’s weight.  

 
 
 
3.  With the parent remaining on the scale press          . The 

scale will zero

, and the word “TARE” will appear at the 

bottom center of the display 

along with the parent’s tared 

weight.  

4. 

Carefully place the child into the parent’s arms. The scale 
will automatically deduct the weight of the parent and 
only display the child’s weight.  

 

 
Note:

 For an a

lternate method, after taring the parent’s 

weight in step 3, have the parent step off of the scale. 
With no weight on the scale, the display will show the 
tare value as a negative amount. Have the parent to step 
back onto the scale while holding the child. The scale will 
automatically deduct the weight of the parent and only 
display the child’s weight.  

 

5. 

The tare value will remain on the display until changed, 
cleared, or until the scale is powered off. To remove the 
Tare, press          . The word “TARE” will disappear from 
the bottom of the display, and the tare value will be 
cleared from the display. The scale will resume normal 
operation.

 

 
 
 
 
           
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Tare 

Tare 

Summary of Contents for 2650KL

Page 1: ...Wheelchair Scales User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM2650KL_2650KG_2700KL_2700KG_E_REV20180625 Equipped with Everlock E...

Page 2: ...ounting Instructions 8 Powering the Scale 10 Set Up Instructions 11 Everlock Instructions 14 Operating Instructions 15 Transport Instructions 21 Maintenance and Cleaning 22 Calibration 23 Troubleshoot...

Page 3: ...AC adapter cord for cracking fraying or for broken bent plug prongs Prior to using this scale ensure that the AC adapter is plugged into an outlet with the rated voltage appropriate for operation Ens...

Page 4: ...0 lb x 0 2 lb 454 kg x 0 1 kg 2650KG 2700KG 454 kg x 0 1 kg Power Requirements 100 240V AC 50 60Hz to 9V DC Adapter Type UE15WCP 9 VDC 500 mA Item ADPT30 included or 6 D batteries not included Environ...

Page 5: ...to other electronic devices please contact Health o meter Professional Scales Technical Support available during normal business hours at 1 800 638 3722 Scale Disposal Health o meter Professional scal...

Page 6: ...ly remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid damaging the scale parts when unpacking do not use a box cutter knife scissors or any...

Page 7: ...level the scale Be certain that the feet are not completely tightened into the platform for proper load cell operation Open the scale to its operational position by unfolding the wheelchair ramps 3 C...

Page 8: ...or placed on a table up to 71 1 8 m from the platform 1 Place the display head with its attached mounting bracket mount on the wall in the desired location Ensure that the display head is level and th...

Page 9: ...1 8 3 2 mm from the wall Warning Extending the screw heads beyond 1 8 3 2 mm can damage the display assembly 6 Obtain the display head and the load cell cable Plug the load cell cable into the load c...

Page 10: ...unscrewing the battery cover screw 2 Remove the battery pack from the display head Replace or install fresh batteries making sure to install them in the proper direction Health o meter Professional re...

Page 11: ...is locked per the instructions above users can still view the alternate weight measuring unit by enabling the Unit View option Enabling the Unit View option allows the user to press the Unit button t...

Page 12: ...time of inactivity before power off 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu p...

Page 13: ...n powered on rather it will use the last zero setting it had from its most recent use There will be no delay in using the scale if the Auto Zero is disabled Health o meter Professional recommends that...

Page 14: ...1 800 815 6615 1 Press to access the Help menu options 2 While in the Help menu press to select Set Options Press to advance to the next menu 3 While in the Set Options sub menu press to scroll throu...

Page 15: ...f a weight value to be deducted from the weight on the platform Also releases tare weight returns display to zero Toggles between pounds and kilograms If weight measuring unit is locked and Unit View...

Page 16: ...ared weight of the object from the gross weight of the patient and object together The display will show the patient s weight 5 With no weight on the scale the display will show the tare value as a ne...

Page 17: ...ent The display will show the patient s weight 5 The weight may be held on the display by pressing The scale will continue to display the patient s weight after they are off the scale and will remain...

Page 18: ...on the display by pressing The scale will continue to display the patient s weight after they step off the scale and will remain on the display until released or the scale is powered off To release p...

Page 19: ...patient remaining on the scale press 3 The scale will prompt entry of the patient s height Use the arrow keys to enter the patient s height Note When weighing in pounds the height is displayed in inch...

Page 20: ...e parent s tared weight 4 Carefully place the child into the parent s arms The scale will automatically deduct the weight of the parent and only display the child s weight Note For an alternate method...

Page 21: ...off of the mounting screws 3 Unplug the load cell cable from the scale platform and the display head Carefully store the load cell cable during transport 4 With the scale laying flat on the floor fold...

Page 22: ...sinfecting Proper care and cleaning is essential to ensure a long life of accurate and effective weighing Caution Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Hea...

Page 23: ...s should be used for calibration Warning Always ensure that the scale is returned to the correct UOM after scale calibration 1 Press to access the Help menu options 2 While in the Help menu press to s...

Page 24: ...ibration weights to be placed on the platform Press after calibration weights are placed on the platform 9 The display will show Calibrating during the calibration process 10 If calibration is success...

Page 25: ...balanced Completely hand tighten the feet then turn them back 1 full turns 4 Scale is out of calibration 4 Check weight with known weight value Weighing is performed but the weighing process takes too...

Page 26: ...This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get Warranty S...

Reviews: