26
CALIBRATION
Calibration
These scales have been factory calibrated, and does not require calibration prior to use.
For maintenance, the scale can be calibrated. Calibration of your 2500KL or 2600KL scale is
performed using kilograms (kg) or pounds (lb), according to the units of measure used upon
entering into the calibration path. The following calibration assumes the scale is set to pounds,
and 400 lb (the default weight) is employed as the calibration weight.
Scale Calibration
Operator Action
Display
1. Make sure the display is turned off, then press and hold the
HOLD/RELEASE
button while pressing the
ON/OFF
button.
“U 01” followed by a 5 digit
‘idle scale count’ number.
2. Press the
ENTER
button.
“CAL”.
3. Make sure there is nothing on the scale, then press the
ENTER
button again.
“Clear” followed by
“0400.0LB”.
4. Use the
►
and
▲
buttons to set the calibration weight to a
value other than the default weight. The calibration weight should
be as high as possible, with 400 lb as a minimum calibration
weight.
Shows “400.0” or entered
calibration weight.
5. Press the
ENTER
button.
Alternating “Load” followed
by “400.0LB”.
6. Load the scale with the required weight.
Display increments showing
raw weight data.
7. Without touching or shaking the scale, allow the display to
stabilize, then press the
ENTER
button.
“SUCC” or “Err1”.
8. If the calibration is successful, the display will show SUCC and
turn off. If the calibration is not successful, the display will show
Err1 and turn off. Contact customer service for an Err1 display
(+1 800-638-3722).
-------
.
Summary of Contents for 2500KL
Page 10: ...2500KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 1 Figure 2 10...
Page 12: ...2500KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 5 Figure 6 12...
Page 13: ...2500KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 7 13...
Page 17: ...2600KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 8 Figure 9 17...
Page 18: ...2600KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 10 Load Cell Port 18...