ASSEMBLY INSTRUCTIONS (CONT.)
Health o meter
®
Professional Scales recommends that assembly be performed by two
people.
Warning: Do not drop scale, as this can permanently damage the load cells.
Caution: Do not insert the batteries into the display head until it is mounted onto the scale.
1. Remove the scale platform/base from
the carton and place on a flat, level, and
dry surface. Locate the handrail and the
load cell cable protruding from the
bottom of the right side. Feed the load
cell cable into the large center hole of
the right handrail mounting hole in the
scale base.
2. Position the handrail over the left and
right mounting holes on the scale base.
Be sure not to pinch the cable. Place the
eight M8 washers (Part D) onto the eight
M8 1.25 x 20mm bolts (Part A) and insert
into the mounting holes on the scale
base. Using the included 5mm Allen
wrench, tighten the bolts.
3. Gently tip the scale forward or onto its
side to access the underside of the
platform.
Caution: Take care to hold the ramp down
when moving the scale - the ramp can flip
open when the scale is tipped forward
.
Grasp the load cell cable protruding from
the bottom of the platform and feed it
through the protective metal shield.
4. Plug the load cell cable into the load cell
port located on the back of the platform.
Pull the excess cable through the top of
the center handrail.