
WE2111
A3983-1.0 en/de/fr
33
"
Use the
and
buttons to enter the weight used,
10 kg in this case.
"
Load the scale with 10 kg and press the
button.
S in P
is displayed as long as the zero balance is
being performed.
10.00
flashes in the display when
the adjustment has been performed.
"
Press the
button to quit the menu.
"
Press the
button for several seconds to save the
settings.
The following are displayed in sequence before the
actual measured value appears:
SAviNG
, firmware
version (e.g.
P60c
) and calibration counter (e.g.
C00008
).