Measuring
FS22 SI
A4249-8.0
HBM: public
49
4.4.5
Fuse Failure
Electrical shorts on the power supply may cause prob
lems to the interrogators. The Industrial BraggMETER SI
on its Rack-Mountable format is directly connected to the
100-240 V power line meaning that it has no external pro
tection. Replaceable fuses are available on the interroga
tor's back-side next to the power connector (
number
5
in
). To replace the fuses, release the top and bot
tom springs of the fuses' support (
remove the piece (
).
Fuse characteristics:
S
Rated voltage: 250 Volt AC
S
Interrupting rate: 2 Ampere
S
Number of fuses: 2
Fig. 4.12