57
FS22DI
BRAGGMONITOR DI USER INTERFACE
Fig. 7.15
Spectral View
The Spectral View can only represent one channel at a time. Channel selection can be
performed on number
9
.
Important
Whenever the spectral view is active, the sampling rate of the interrogator is set to 50 S/s,
reverting back to the previously operating sampling rate when another tab is selected.
The number of FBGs identified in the selected channel is represented in number
10
in
.
Information
The number of detected peaks per channel is limited in accordance to the sampling rate. In
case there are more peaks than the limit, the lower wavelengths will be eliminated from the
list. For more detailed information about the maximum number of sensors per optical
channel at a specified sampling rate please refer to the product Datasheet.
Measurement Configuration
Measurement settings can influence sensor readings. Gain and Threshold values should
be defined according to the FBG sensing network status, so that peak amplitude is