![HBK FS22DI User Manual Download Page 56](http://html1.mh-extra.com/html/hbk/fs22di/fs22di_user-manual_2111693056.webp)
FS22DI
BRAGGMONITOR DI USER INTERFACE
56
Fig. 7.14
Numerical View
7.2.3
Spectral View
The Spectral View tab is only accessible when data acquisition is stopped.
This tab represents the spectral response of the sensors connected to the interrogator.
The full spectrum contains approx. 7020 data points corresponding to the reflected opti
cal power in dBm from 1500 nm to 1600 nm with an approx. 15 pm sampling resolution
(number
1
in
). For each detected FBG peak the pair of wavelength and power
values are represented in a table format (number
11
in
)