10
The use of BCY01 accelerator is recommended under the following conditions:
S
Humidity is too low (rel. humidity
t
30 to 40%)
S
Surfaces react acidically
S
Temperatures are too low
Important
Only use the BCY01 accelerator when one of the above-mentioned conditions is present.
Improper application can lead to shock curing. The stresses then present in the adhesive
layer will lead to unstable measuring points and will reduce the maximum elongation of the
joint.
Tip
Practical experience with CA80 shows that curing difficulties are mainly caused by
applying the adhesive too thickly.
8
STORAGE
Store the bottles standing where possible.
Protect the superglue CA80 against heat, sunlight and moisture, store it cool and dry.
When stored correctly, usable at least until the best-before date indicated on the packag
ing.
9
SPECIFICATIONS
Temperature limits
For zero-point related measurements
-55 °C … +100 °C
For non zero-point related measurements
-70 °C … +120 °C
Solubility
Dimethyl sulphoxide, alkali, boiling water
Swelling
If stored for long periods in esters
(ethylacetate) and ketones (acetone,
methylethylketone)
Summary of Contents for CA80
Page 1: ...CA80 ENGLISH DEUTSCH FRAN AIS Instructions for use Gebrauchsanweisung Instructions d emploi...
Page 3: ...CA80 ENGLISH DEUTSCH FRAN AIS Instructions for use...
Page 13: ...CA80 ENGLISH DEUTSCH FRAN AIS Gebrauchsanweisung...
Page 25: ...CA80 ENGLISH DEUTSCH FRAN AIS Instructions d emploi...
Page 36: ...HBK Hottinger Br el Kjaer www hbkworld com info hbkworld com A05903 01 Y00 00 7 0104 0036...