Assembly Instructions M8103-EN 

 

M8103-EN   2020-01 

 

3/4

 

 

  

 

8.

 

Replace removed cable ties or replace with new ones. 

  

9.

 

Close and secure the front panel. 

 

10.

  Reconnect mains plug and switch on main switch. 

 

 

11.

  Check that the unit is free of pressure  

(

 Pressure gauge rear of unit). 

 

 

12.

  Calibrate the pressure sensor 

   

 

with pressure sensor system (BP 1) 

  

 

Fig. 1: Calibrate the pressure sensor 1 

  Read off the current 

System pressure actual value

 under 

Display \ Actual value

  Set the parameter 

Pressure sensor 1 offset

 under 

Service \ 

Calibrating \ Pressure

 according to the following calculation: 

 

Pressure sensor 1 offset

 new = 

Pressure sensor 1 offset

  

current 

– 

System pressure actual value 

current

  

     

 

with pressure sensor main line (BP 2) 

  

 

Fig. 2: Calibrate the pressure sensor 2 

  Read off the current 

Main line pressure

 under 

Display \ 

Actual value 

  Set the parameter 

Pressure sensor 2 offset

 under 

Service \ 

Calibrating \ Pressure

 according to the following calculation: 

  

Pressure Sensor 2 offset

 new = 

Pressure Sensor 2 offset

 

current 

– 

Main line pressure

 current

 

   

 

13.

  Check calibration pressure sensor 

  Parameter 

System pressure actual value

 resp. 

Main line 

pressure

 under 

Display \ Actual value

 must be 0 bar ± 0,1. 

  If this is not the case, item calibration pressure sensor must be 

repeated. 

 

 

ATTENTION! 

Risk of incorrect calibration! 

False calibrations can lead to faults witch the unit. 

Therefore: 
–   Check the calibrations. 

  

 

14.

  Press the 

 key to switch on the unit, check unit functions 

and check it for leaks. 

 

 

15.

  Switch the unit off by press the 

 key. 