HB Products HBPH-MK2 Instruction Manual Download Page 7

 

  

 

Instructional manual- HBPH-MK2 

– UK-004 

  

  

Page 7 

 

Replacement of Salt Bridge   

1.

 

If the system cannot be calibrated after cleaning the probe, it may be necessary to replace the 
standard cell solution and/or the salt bridge. A salt bridge kit is available from HB Products for this 
purpose.   

2.

 

Hold the probe vertically electrodes up. Remove the used salt bridge. The salt bridge is a 
hexagonal-

shaped capsule that can be removed using a #3 Phillips screwdriver. ½” wrench or pliers 

can also be used.  

3.

 

Discard the used salt bridge.  

4.

 

Dispose of the used solution inside the bridge chamber and rinse with pH 7 solution or distilled 
water.  

5.

 

Refill the chamber with a reference solution supplied with the salt bridge, up to the tip of the 
electrode inside the chamber. DO NOT OVERFILL. It is important to leave space for the salt bridge 
thread and a small amount of air.  

6.

 

Screw the new salt bridge into the cavity until finger tight.  If needed, use a #3 Phillips screwdriver 
to make sure that the salt bridge edges are aligned with the front of the probe face.  

  

 

Troubleshooting  

A sensor that is not functioning properly will output a current that is out of range of the specifications 
listed in Checking Sensor section. 

The change in output with calibration standard constitutes the span. For a pH probe, the span between 
pH 4 and pH 7 or between pH 7 and pH 10 should be between 2.75 and 3.5 mA.  

Summary of Contents for HBPH-MK2

Page 1: ...on 3 Cold Liquid applications 3 Electrical connection 4 Storage 4 Maintenance and calibration 4 Calibrating the Sensor 5 Checking the Sensor 5 Probe Cleaning 6 Replacement of Salt Bridge 7 Troubleshoo...

Page 2: ...period thoroughly clean the sensor put a piece of cotton ball with a few drops of water into the protective cap and place it on the sensor This keeps the junction from drying out which causes slow re...

Page 3: ...duit Do not run the power cable or control cables in the same conduit with the probe interconnect cable 3 Remove the protective plastic cap from the end of the probe before placing in service 4 The se...

Page 4: ...measurement If you use the sensor for a precise measurement of the pH value you need frequent checks of the calibration The drift of the sensor is maximum 0 03pH per day but that is the maximum figure...

Page 5: ...Allow the temperature of the buffer and probe to stabilize at room temperature 4 Write down the reading on the multimeter as mApH7 in the example called mApH7 The reading should be between 11 and 13 m...

Page 6: ...ied the probe is defective or the salt bridge needs to be replaced 8 If the span of the probe drops below 2 75mA than the probe still can be used Adjustments will have to be made to the receiving devi...

Page 7: ...th pH 7 solution or distilled water 5 Refill the chamber with a reference solution supplied with the salt bridge up to the tip of the electrode inside the chamber DO NOT OVERFILL It is important to le...

Page 8: ...iled Sensor is non repairable The output is outside of 4 20mA The sensor electronic has failed Sensor is non repairable Parts and Accessories Description Part 3 pack salt bridge replacement kit HBPH S...

Reviews: