Manual Conpreva
Set the current
You reach the current setting, if you selected a pro-
gram with current (ES, BF+ES, Trigger). The
effective current is represented on the one hand
numerically, on the other hand on a logarithmically
scaled animation.
lock is open:
With the
turn switch
you can adjust the amperage.
The feeling should not be painful.
lock is locked:
Turn the
turn switch
against the clockwise direction.
The lock opens and deal as described above.
If the amperage does not increase inspite of an open
lock, control the contact of the electrodes. The device
refers you with an animation to check it.
The feeling of the current heavily depends to the
adjusted frequency and pulse width.
The current setting is skipped, if your physician set a
fixed current value. You can always reduce or
increase the current during the exercise.
Exercise with
electrical
stimulation uses
only channel 1 for
measurement, so
only one emg scale
is shown
Course of exercise
Course of exercise
Flowing current in
mA and graphical
EMG-setting
Depending to the selected EMG mode, (you recognize
this by the M, S or A in the left lower corner) proceed
please as follows:
• M - manually EMG-setting
To set your individual EMG range tense up as
strongly as possible the concerned musculature.
Now turn the
rotary switch
in such a way that the
deflection on the EMG scale is approximately 3/4
(see illustration below)
• S - semi automatic EMG-setting
Tense up the concerned musculature, until the
EMG range does not change any longer. Usually
the setting is finished after 4 seconds
• A - automatic EMG-setting
If you choose the automatic EMG setting, the
optimal EMG range will be adapted during the
exercise. The conventional EMG setting is
skipped.
EMG-scale