Step 2

Before doing any assembly, pre-test motors. Plug power cord into grounded wall outlet. The electrical rating for Series 
1440 is 110V AC, 60Hz, 1.6 Amps. The 3-prong hospital grade grounded plug must be plugged in 3-pronged, 
grounded, non-isolated, correctly polarized 110V AC receptacle. 

Failure to do so could result in personal 

injury or damage to equipment.

A. Step on UP button and elevate unit about 5” - 6”. 
B. Step on DOWN button and lower down as far as it will go. 
C. Step on UP button and elevate up to maximum height.

NOTE: 

During the pre-testing, the green light on the Control Box should be on to indicate there is power to the control 

box. If for any reason there is no response, see the trouble shooting guide on page 8 for possible problems - most 
common of which is loose plug connections into control box.

You have completed initial test. Unplug power cord from power source as electrical power is not needed for remain-
der of assembly.

110V Power 
Cord

Detail of Control Box

Green Light

Foot Control Cord and Plug

Underside of Control Box

Port for 
Actuator A1

Port for 
Actuator A2

110 V Power 
Cord

3

Summary of Contents for Powermatic 1440 Series

Page 1: ......

Page 2: ...location where you want to install the mat platform X Distance between base plates varies as follows Size of Top Distance Between Plates 6 x 8 52 5 x 7 4 x 7 45 4 x 6 39 3 8 A1 Actuator Wire Hard Con...

Page 3: ...on and lower down as far as it will go C Step on UP button and elevate up to maximum height NOTE During the pre testing the green light on the Control Box should be on to indicate there is power to th...

Page 4: ...3 holes in black angle irons line up with the 8 pre drilled holes in underside Secure each black angle iron to subtop with 4 1 4 20 x 7 8 machine screws D into pre threaded t nuts in underside of top...

Page 5: ...ol box to underside Secure power cord and footswitch wire together to underside of top with two cable ties as shown Step 7 Use cable zip ties to properly secure actuator wire to underside of top First...

Page 6: ...must rest on base plate and butt t to black box around actuator Secure shelf with 3 1 3 4 x 8 at head phillips head screws per each side total of six screws needed Apply pressure as holes are not pre...

Page 7: ......

Page 8: ...into wall outlet If unit still does not work verify that the wall outlet is active by plugging in another appliance or lamp Could be shut down due to ground fault de activation or faulty outlet 2 Ver...

Page 9: ...peci cations Input 120 VAC 1ph 60 Hz 2 5 Amps Max Output 24 VDC 6 Amps Environmental Conditions Normal Operations Ambient Temperature 10 to 40 C 50 to 104 F Air Humidity Max 85 Environment Conditions...

Page 10: ...PLUG FROM CONTROL BOX The two lines may only be detached from the control box by means of special plastic wrench with which the brackets can be separated by pulling the plug see illustrations The bra...

Reviews: