4    Install 

 

11 

 

 

6.  Connect the discharging 

power pack to the supply 
voltage. Always use the 
supplied IEC power cord 
with its integrated ground 
wire. 
•  Plug in the IEC power 

cord into the cold device connection of the discharging power 
pack. 

•  Plug in the mains plug only to a socket with functioning 

ground conductor. 

 

 

 

Contact and separation spark-overs! 

When the ionizing unit is plugged in or unplugged while the 
discharging power pack is switched on, spark-overs will occur at the 
HV connection. This may result in defects in the discharging power 
pack. 
•  Switch off discharging power pack before plugging in/unplugging 

ionizing unit. 

 

 

7.  Connect the ionizing unit to 

the HV terminal of the 
discharging power pack. 
•  Insert the ionizing unit’s 

HV plug in the HV 
terminal of the 
discharging power pack 

and press the HV cable 
until it reaches the stop. 

•  Screw the screw cap 

onto the HV terminal and tighten by hand. 

 

 

NOTE: 

Protect unused HV terminals against the ingress of environmental 
substances using the blind plugs. The blind plugs must be clean, 

dry and all grease must be removed. 

 

 

8.  The discharging power pack is ready for operation. 
 

 

 

 

 

 

Summary of Contents for METTLER TOLEDO EN Series

Page 1: ...Static Line Operating instructions Discharging power pack EN SL MT Ident number 01 7907 055 100 V 01 7907 005 115 V 01 7906 005 230 V Keep for future use GB...

Page 2: ......

Page 3: ...2 Safety 6 2 1 Intended use 7 3 Product overview 8 4 Install 9 5 Operate 12 6 Troubleshooting 13 6 1 Replacing fuse 14 7 Accessories spare parts 15 8 Technical data 16 8 1 Characteristics and specific...

Page 4: ...eviated HV in these operating instructions e g HV terminal The illustrations in this document are a simplified representation of the product They render only the technical facts and provide support fo...

Page 5: ...ose of with household garbage Caution danger spot warning 1 2 Symbols on the discharging power pack WARNING High voltage ATTENTION Only plug in unplug the ionizing unit at the HV terminal when the dis...

Page 6: ...In the event of visible damage or suspected electrical defects take the discharging power pack out of operation immediately and secure against inadvertent reuse Protect the discharging power pack from...

Page 7: ...tmospheres The discharging power pack is intended exclusively for the supply of alternating high voltage to HAUG ionizing units with X 2000 connector In combination with an ionizing unit electrostatic...

Page 8: ...8 3 Product overview A Mains switch lights up green when discharging power pack is switched on B Ground connection terminal C 2 x HV terminal D Fuse holder with fuse Replacing fuse refer page 14 E Col...

Page 9: ...ontinuous overloading of the discharging power pack may result in failures Never exceed the permissible connected length Never install the discharging power pack on a surface generating or radiating h...

Page 10: ...position in which it is installed We recommend installing the discharging power pack with the HV terminals pointing downwards to protect them from moisture oil and dirt 4 Ensure that the discharging...

Page 11: ...l occur at the HV connection This may result in defects in the discharging power pack Switch off discharging power pack before plugging in unplugging ionizing unit 7 Connect the ionizing unit to the H...

Page 12: ...pack and the ionizing unit are connected and installed as specified in the operator instructions 1 Switch on the discharging power pack using the mains switch A The mains switch will illuminate green...

Page 13: ...ing unit for checking to HAUG GmbH Co KG the address is provided on the back of the envelope Fault Cause Troubleshooting Mains switch does not illuminate Power failure Inspect the power fuse Defective...

Page 14: ...red fuses Never bridge the fuse The unit type and the rated voltage are indicated on the nameplate 1 Disconnect discharging power pack from supply 2 Determine and remove the cause for the blown fuse 3...

Page 15: ...e authorized local sales agent of Mettler Toledo Article Order number Point electrode 11107765 Optional second electrode 11107762 U Electrode universal 11107764 U Electrode small 11140161 Ionizing kit...

Page 16: ...23 C HV terminals 2 High voltage 6 7 1 kV Short circuit current Ik 5 mA Cannot be used in pulsed mode 8 2 Supply voltage Unit type Nominal value Frequency range Power input 01 7907 055 100 V 10 50 60...

Page 17: ...e mean seal level to 2000 m Temperature Rated application range 5 to 45 C Limit range for storage and transport 15 to 60 C Relative humidity Rated application range A RF B C Air pressure Rated applica...

Page 18: ...l data 18 8 4 Housing Protection type IP 20 Protection class I Overvoltage category II Pollution severity 2 Mains connection About IEC power cord Dimensions Height 170 mm Width 110 mm Depth 100 mm Wei...

Page 19: ...t from the HV terminal 4 Remove the discharging power pack from the production process 9 1 Storing Always store our products in a dry and cool place 9 2 Disposing Never dispose of electrical appliance...

Page 20: ......

Page 21: ...NOTES...

Page 22: ...NOTES...

Page 23: ......

Page 24: ...EN SL MT v09gb docx V09 22 01 2019 D 0282 GB made by...

Reviews: