background image

 

 

PREPARING ITEMS FOR STERILIZATION

 

Because of the possibility of transmitting micro-organisms from patients to professional 
personnel, contaminated instruments should be handled carefully. Hand-Scrubbing of sharp 
instruments should be avoided where possible and an ultrasonic cleaning device employed for 
this purpose. 
 
All Instruments should be cleaned of blood, debris or extraneous material before being placed in 
the instrument tray. Upon completion of the cleaning process, instruments must be thoroughly 
rinsed in cold running water to remove any residue of ultrasonic solution or soap and then towel 
dried. 
 
INCOMPLETE RINSING MAY RESULT IN THE ACCUMULATION OF SOAP ASH AND 
FOREIGN DEBRIS IN THE INSTRUMENT TRAY OR STERILIZATION CHAMBER.  
 
Small objects such as burs and needles or items difficult to towel dry should be dipped in Vapo-
Steril after rinsing to eliminate water. Handpieces should be cleaned according to 
manufacturer's instructions before sterilization. Lubricate handpieces according to 
manufacturer's instructions 

following

 the sterilization cycle.  

 
Use only clean, preferably sterile oil. Handpieces may be placed in the tray with other 
instruments. If storage of sterile instruments is desired, sterilize in Harvey Sterilization Indicator 
Bags. These bags permit penetration by the chemical vapor, but preclude contamination by air-
borne bacterin.  
 
AVOID HEAVY OR MULTIPLE TIGHTLY WOVEN WRAPPINGS OR SEALED CONTAINERS. 
FOR STERILIZATION, CONTENTS MUST BE EXPOSED TO THE CHEMICAL VAPOR.  

 

STERILIZING

 

Line the instrument tray with a Harvey Sterilizer Tray Liner and place items thereon. Close and 
latch the door. Turn the control knob (see figure 1 below) to "Pressurize". It is not necessary tp 
turn th "Time" control since the 20 minute cycle time is set automatically. The automatic cycle 
may be overridden by turning the "Time" control to other than 20 minutes.

 

 
The sterilization cycle is automatically controlled. Timing for 20 minutes will not begin until the 
chamber is pressurized to 20 psi. This automatic feature insures that each chamber load, 
regardless of size, will be exposed to the chemical vapors for at least 20 minutes at optimum 
pressure and temperature. 
 
Because of the chemical action of Vapo-Steril, the sterilization cycle begins upon admission of 
solution into the heated chamber. The sterilization cycle consists of two phases: The 
condensation phase (0-20 pounds pressure) and the vaporization phase (20-40 pounds pressure). 
During the condensation phase, pressure rise time will vary depending on the number and size of 
items being sterilized.  

Reviews:

Related manuals for CHEMICLAVE 5000