TROUBLESHOOTING
Some of the cells
appear to be deflated.
This is normal for alternating pressure therapy
.
The mattress is made up of individual air cells. The alternating section is split up into 2 sections consisting of odd
cells e.g. 1,3,5 etc and even cells e.g. 2,4,6 etc. These two sections will alternate through a 10 minute cycle in
which time both sets of alternating air cells will have inflated and deflated sequentially
.
CHECK THA
T
FAUL
T
ST
AGE 2 CHECK
The system does
not appear to be
alternating.
Check that the static
mode is not turned on.
Carefully mark one of the
inflated cells with a pen.
Ensure that there are no
kinks in the pipework down
the side of the mattress.
1.
2.
3.
Monitor the cell for 7 minutes
to see if it deflates.
Straighten out any kinked
pipes.
1.
2.
IF PROBLEM PERSISTS
Contact Har
vest Healthcare technical
support.
Before calling:
Please ensure you have the serial number
and model of equipment.
Please record details of the results of the
recommended tests. (Notes pages are
provided at the back of this user manual).