CLEANING & CARE
PUMPS
For general cleaning wipe with a soft cloth dampened with a mild detergent and water
solution. This may be followed by either wiping with a sodium hypochlorite solution to a
dilution of 1000ppm (parts per million) or by using alcohol wipes.
DO NOT USE HYPERCARBONATE, PHENOL-BASED CLEANING
SOLUTIONS, ABRASIVE COMPOUNDS OR CLEANING PADS.
NOTES
Following the use of a detergent and or disinfectant solution rinse the mattress cover
with clean water using a clean cloth and allow to dry.
Frequent or prolonged exposure to high concentrations of disinfectant solutions will
reduce the useful life of the mattress cover.
Where high concentration disinfectants e.g. > 10,000ppm chlorine releasing
agent (e.g. Haztab or bleach) or combined cleaning/chlorine releasing agent (e.g.
Chlorcleam, Actichlor) and detergent solutions are used to remove blood or other
body fluids, mattresses should be thoroughly rinsed with clean water to remove any
residues. This will help prevent any long term compatibility issues associated with
disinfectant residues.
Alternatively, disinfection may be achieved by laundering at temperatures not
exceeding 80°C for 10 minutes which may include a chlorine rinse.
•
•
•
•
TRANSPORT & STORAGE
Storage conditions as follows:
–15 °C without relative humidity control; and +40 °C at a relative humidity up
to 93%, non-condensing. An atmospheric pressure range of 700 hPa to 1 060 hPa.
Suitable for all standard modes of transport when in the correct packaging.
Operation Conditions:
A temperature range of +5 °C to +35 °C; A relative humidity range of 15% to 93%, non-
condensing; and Operational Atmospheric Pressure: 700 hPa to 1060 hPa. Suitable for
pollution degree 2. Operational altitude ≤ 2 000 m.