background image

TECHNICAL INFORMATION
Membrane

Material

Polyethersulfone (PES)

Molecular Cut-Off

500kDa, 2MDa

Outer Diameter

0.28 mm

Length

1, 2, 3 and 4 mm

Probe Shaft

Material

Stainless-steel

 

Diameter

0.40 mm

Length

14 mm

Internal Volume

Inlet Volume

0.1 µL

Outlet Volume

1 µL

CMA 11 

55

 kDa

 

2MDa

 

Microdialysis Probe User’s Manual

0.40

0.2

8

shaft

steel shaft

outlet

intlet

membrane

Instructions for CMA 

11 5

00

 kDa

 & 2MDa

 

Microdialysis Probe

s

1.

2.
3.

4.

5.
6.

7.

8.

9.

10.

Note: The 500 kDa 

and 2 MDa 

Cut Off membrane has to be use in a Push-Pull system to avoid loss of fluid out to the 

surrounding tissue (ultra-filtration). Here the use of a CMA syringe pump and REGLO ICC Pump is recommended.

Load the peristaltic pump with the "FEP Tubing Connector Peristaltic 

Kit

". Adjust the lengths of the FEP tubing if needed for a 

lower dead volume; recalculate the outlet volume if needed (1.2 µl/100 mm length)

Run the pump to make sure that liquid leaves the tip of the syringe cannula.

Fill a microsyringe with perfusion fluid and mount it in the CMA Syringe Pump. The Perfusion Fluid must be clean, at 

room temperature and preferably degassed.

Place the inlet end of the “FEP Tubing Connector Peristaltic 

Kit

” in a beaker with Perfusion Fluid and flush the pump to fill all 

tubing with perfusion fluid. Make sure there is no air bubbles in the complete tubing set.

S

et the pump to the required perfusion flow, usually 1 – 5 µL/min.

Prepare a desired length of tubing with a Tubing Adapter on both ends and connect the tubing to the inlet of the probe. 

Remove the protection tube from the probe carefully. Short cannula = inlet, long cannula = outlet. Tubing Adapters and FEP 

tubing should be used for all connections. T

o facilitate the handling of Tubing Adapters, they should be soaked in Ethanol 

70% for minimum 10 minutes.

Mount the probe in a Probe/Guide Clip on the CMA 130 In Vitro Stand. Put the probe membrane into a vial filled with 

perfusion fluid. Connect the inlet tubing of the probe to the syringe cannula by sliding the Tubing Adapter over the cannula.
Flush the probe with perfusion fluid at 8-10µL/min for 4-5 min to wash out air. Knock on the shaft of the clip to help the air to 

flush away. At this process the Ultra High Cut Off membrane will look as leaking but this is due to ultra-filtration of fluid 

through the membrane.

Set the pump to the required perfusion flow, usually 1 – 5 µL/min. Make sure both pumps have the same flow rate.

Connect the inlet “FEP Tubing Connector Peristaltic 

Kit

” at the peristaltic pump to the outlet of the probe

11

.

12

.

13

.

Lift up the probe from the vial and check that the membrane doesn’t ultrafiltrate or dry out; it should look filled out but 

not sweating. Lower the probe into the vial and control the outlet volume for the complete system.
The system with the probe is now ready for use.

When changing sample vials, remember to consider the internal volume in the system (see TECHNICAL INFORMATION). 

This causes a delay that must be calculated when using low perfusion rates and short sampling times

Reviews: