background image

AgilePulse PLUS

Quick Start Guide

BTX

 

84 October Hill Rd • Holliston, MA 01746, USA

phone: 1.508.893.8999  •  fax: 1.800.429.5732 

web: www.btxonline.com  •  email: [email protected]

Publication 5423-031 Rev 1.0

Protocol Setup

Pulse protocols can be viewed and/or modified on the Protocol screen 
which is accessed from the Welcome screen: 

 Touch 

SETUP 

 Touch 

PROTOCOL 

Parameters for up to five pulse groups are set by selecting the 
appropriate box and scrolling through the available values. Subsequent 
groups are activated by selecting at least one pulse in the number field 
and are conversely deactivated by selecting zero pulses in the number 
field. The adjusted protocol is saved to internal memory when the 
HOME button is touched.

• 

Amplitude defines voltage of the pulses.  

• 

Duration defines the length of the pulses. 

• 

Pulse Interval defines the inter-group interval (pulse interval time 
within a single group).

• 

Group Interval defines the intra-group interval (pulse interval 
time between one group and the next group. Group Interval also 
defines the interval in the middle of a group when polarity flip 
settings are turned on.

• 

Number defines the number of pulses.

• 

Polarity defines the directionality of the electrical current flow 
during the pulses. Options include Positive, Negative, Flip 
- → +, and Flip + → - . For the Flip polarity options, the first half 
the pulses of the group will be delivered in the one polarity, then 
the second half of the pulses of the group will be delivered in the 
opposite polarity. For Flip polarity groups with an odd number 
of pulses, the extra pulse will be delivered prior to the Flip. For 
example, a Flip + → - group with pulse number set to 3 will have 2 
positive polarity pulses followed by one negative polarity pulse

Pulse Delivery Procedure

1.  Prepare sample for electroporation with BTX Large Volume 

Chamber, Flatpack Chamber, cuvette or in vivo electrode.

2.  Make proper connections between generator, electrode, and 

sample. 

3.  Enter Cell Line, Transfectant, and Chamber data on the Welcome 

screen, then tap Next to advance to the Pulse Delivery screen.

4.  Touch the Load icon. The system will automatically deliver a short 

low voltage pulse to estimate the load resistance. If the resistance 
is too low, the pulse will not be delivered. If the resistance is 
within acceptable limits, the Ready button will turn yellow.

5.  Touch the Ready button. When the system is charged, the Start 

button will turn yellow. 

6.  Touch the Start button. The system will deliver the set pulse 

protocol. When delivery is complete, the Done button will turn 
green and the pulse log will be displayed in the text box. 

7.  Touch the Done button to return to the Welcome screen.  

The identification fields will be reset. Repeat the procedure for 
each pulse delivery. 

Protocol Screen

Pulse Delivery Screen—Pre-Pulse

Pulse Delivery Screen—Post-Pulse

BTX AgilePulse PLUS Quick Start Guide    •    www.btxonline.com