CMA 31 Linear Microdialysis Probe User’s Manual

TECHNICAL INFORMATION
Membrane

Material

Polyethersulfone (PES)

Molecular Cut-Off

55,000 Daltons

Outer Diameter

0.26 mm

Length

10 mm

Probe Shaft

Material

Polyimide

Length Inlet Tubing

350 mm

Length Outlet Tubing

100 mm

Total Length ( membrane)

460 mm

Outer Diameter

0.19 mm

Double Tubing Outer Diameter

0.63 mm

Double Tubing Length

12 mm

The probe can be sterilized in its inner package with ethylene oxide.

Cutaneous Implantation, Anesthetized Animal

For CMA 31 Linear Microdialysis Probe: To avoid ultra filtration over the membrane 
the probe should be used with a Perfusion Fluid consisting of a Ringer solution 
containing 30g Dextran 60/1000mL.

1.

Anesthetize the animal. Be sure to keep the animal’s body temperature normal 
during the surgical procedure and the study period by using a CMA 450 
Temperature Controller.

2.

Clean and shave the animal.

3.

Mark the skin with a felt tip marker where the introducer needle should go in 
and exit, leaving at least an 18 mm window if drug transport is studied. Tunnel 
the introducing needle at the appropriate depth required for the experiment.

4.

Load the CMA pump with a Perfusion Fluid-filled syringe (either a Luer lock 
single use syringe or a glass syringe with a fixed needle). Make sure there are 
no air bubbles.

5.

Connect the inlet Luer lock connector to the single use syringe or cut off the 
Luer lock connector from the probe inlet with a sharp cut, and connect it to 
the glass syringe needle via a 

blue

 Tubing Adapter.

6.

Start the pump at 5 µL/min. Confirm that liquid is flowing from the outlet of 
the probe. Maintain this flushing rate for approximately 2-3 min to remove air 
bubbles.

7.

Reduce the flow rate to operational rates, usually 1-3 µL/min.

8.

Remove the protective tubing from the probe. Slide the probe outlet tubing 
through the needle end of the introducer carefully so the membrane sits in the 
middle of the introducer. Make sure not to kink the tubing.

9.

Check that the flow rate is the same as before the insertion. Hold the probe in 
place against the skin while carefully withdrawing the introducer needle.

10. Secure the probe position with surgical tape or surgical glue.

11. Wait at least 30 min for stabilization before starting the sample collection.