1

SUBJECT

PAGE

Warranty and Repair Information  ........................2

General Information ..............................................3

Description ........................................................3

Specifications ....................................................3

Operation ............................................................4-6

Volume Adjustment ........................................4-5

Recommended Setup ........................................6

Dual Rate Compensation Chart ............................7

Ventilation Graph ..................................................8

Maintenance Instructions......................................9

Model 613 Parts and Assembly Drawing ..........10

Parts List ..............................................................11

Repair Kit Information ........................................12

Table of Contents

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for 55-0715

Page 1: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Page 2: ...lator 115 VAC 60 Hz55 0715 Large Animal Volume Controlled Ventilator 230 VAC 50 Hz55 0723 Large Animal Volume Controlled Ventilator User s Manual Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 3: ...Specifications 3 Operation 4 6 Volume Adjustment 4 5 Recommended Setup 6 Dual Rate Compensation Chart 7 Ventilation Graph 8 Maintenance Instructions 9 Model 613 Parts and Assembly Drawing 10 Parts List 11 Repair Kit Information 12 Table of Contents Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 4: ...s limitation on an implied warranty so the above limitation may not apply to you If a defect arises within the one year warranty period promptly contact Harvard Apparatus Inc 84 October Hill Road Holliston Massachusetts 01746 1371 using our U S only toll free number 1 800 272 2775 or dial 508 893 8999 Goods will not be accepted for return unless an RMA returned materials authorization number has b...

Page 5: ...ecifications Volume Adjustable from 30 to 750 cc stroke Note Volume adjust ment is made on crank and therefore accurate volume adjustments cannot be made while ventilator is running Rate Adjustable from 7 to 50 strokes min while ventilator is run ning Phase Adjustable from 25 to 50 of single stroke cycle continuous ly variable while pump is in operation Port Size ID 12 7 mm 1 2 in OD 15 9 mm 5 8 i...

Page 6: ...0 10 40 25 5 35 45 INSPIRATION BREATHS MIN ON OFF 2 0A S B 2 5 CAUTION This pump is not to be used in an explosive atmosphere or with any explosive gases such as ether cyclopropane etc Figure 1 Control Panel for Large Animal Volume Controlled Ventilator Model 613 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 7: ...ke Rate Control is continuously variable while the pump is in operation Stroke Volume Adjustment Stroke Volume is determined by the position of the Connecting Rod on the crank Loosen theAdjusting Nut set the Connecting Rod to the desired position according to the Calibration Panel and then tighten See Ventilation Graph pg 8 CAUTION Stroke volume should not be adjusted while the pump is in operatio...

Page 8: ...he animal to the room atmosphere A Y Connector should be used to couple the upper and lower port tubes to the tracheal cannula The Y Connector should be placed as close as possible to the animal in order to minimize the common pathway between inspired and expired air Tracheal Cannula or Endotrach Tube Y Connector close to animal From Animal To Animal Expired Air Plastic Tubing Air Gas Input Cylind...

Page 9: ...dual phase motor control pump As observed actual stroke rate varies considerably as the phase ratio reaches its outer limits Users should be aware of this relationship and develop appropriate curves to assure experimental data Figure 4 Dual Phase Rate Compensation Chart Phase Ratio Inspiration Typical plot of actual stroke rate at various ratios and strokes Artisan Technology Group Quality Instrum...

Page 10: ...6 20 30 40 100 150 200 300 400 40 50 60 80 15 20 10 30 8 4 5 6 2 3 1 0 5 0 6 0 8 0 4 500 40 Grams Tidal Volume Kilograms Tidal Volume ml at Ambient Temperature Ventilation Graph Tidal Volume vs Weight and Rate for Laboratory Mammals in Resting State Apparatus dead spade must be added Tidal Volume vs Weight and Rate for Laboratory Mammals in Resting State apparatus dead space must be added Figure 5...

Page 11: ...ould be used for lubrication Gear Reducers Reducer housings have no sight gauge or level plug The housings are filled to 3 4 capacity with Hodson 2 1000 No Melt Grease This will be sufficient for a 5 year peri od To relubricate types N 10R6 reducers remove the gear housing clean out the old grease and refill to 3 4 capacity with Hodson Lubricant mentioned above To relubri cate types N 1RH N 3R N 3...

Page 12: ... 023 0606 030 5112 105 5110 003 0606 114 5103 021 0606 041 0606 040 0606 091 0613 023 0606 044 5013 002 5009 087 0606 079 0606 039 0606 031 0606 010 Supplied in Repair Kit 616 Figure 6 Large Animal Volume Controlled Ventilator Model 613 Parts and Assembly Drawing Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 13: ...ntake Valve 1 0606 083 Assembly Exhaust Valve 1 0606 091 Pin Crank 1 0606 114 Volume Adjusting Knob 1 0613 023 Crank 1 0614 030 Male Valve Cover 2 5009 087 Spring 1 5013 002 O Ring Parker 2 111 3 5013 017 O Ring Parker 2 8 1 5013 023 O Ring Parker 2 336 2 5013 038 O Ring Parker 2 21 4 5100 013 Thumb Screw 10 32 1 5103 021 Acorn Nut 1 4 28 1 5110 003 Curved Spring Washer 1 5112 105 Micro Switch Onl...

Page 14: ...on of rings valves springs gaskets and lubricants Also included is a set of instructions for overhauling procedures The kit contains the following items Model No Description Quantity Per Unit 0606 028 Gaskets Intake and Exhaust Valves 4 0606 030 Nylon Plunger Tip 1 0606 060 Oil 3 4 oz Bottle 1 0606 082 Intake Valve Assembly 1 0606 083 Exhaust Valve Assembly 1 0606 084 Silicone Grease 1 5009 152 Sp...

Page 15: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...