Foot
Hardy Floor Scales
11
FIG. 9 ADJUSTING THE FEET FOR LEVEL
Step 5.
Place the spirit level on the platform sur-
face again to check the adjustment for
level. Keep adjusting the legs until the
platform is level and all four leveling feet
are firmly on the floor surface.
Step 6.
Check the distance from the top of the
platform to the floor surface at each cor-
ner. All the measurements should be
within 3º of each other and the platform
should be level.
Step 7.
If the installation requires bolt down plates
to keep the floor scale in place go to the
Optional Bolt Down Plates Section for
instructions.
CALIBRATION
Pre-Calibration
Procedures
Step 1.
Open the Junction box cover. (See Fig. 11
for Instructions)
Step 2.
Thread the C2
®
Certified cable through
the grommet on the side of the scale that
houses the junction box.