172
Chapter
9
•
•
•
•
•
•
INTEGRATED TECHNICIAN
INTEGRATED TECHNICIAN
®
(IT) is an optional diagnostics utility that enables the
operator to rapidly troubleshoot individual load cells. The option requires an HI 215IT or
HI 6010IT Summing Box (shown above) that provides distinct inputs for each load cell.
Without the HI 215IT or HI 6010 Summing Box, it is difficult to isolate the signals from
different load cells. If any load cell fails, the test will output a FAIL response without
identifying the problem load cell. For numerical values, the system will return an average
of all the load cell responses and, in some cases, will return values that cannot be used.
The HI 215IT or HI 6010 Summing Box with the IT firmware can provide the HI 6500 both
average numerical values and values specific to each load cell, including PASS/FAIL
values for each load cell, as shown below.
To view this screen on your system, see
Diagnostics
The number of load sensor(s) is based
on the connections in the IT junction box. Check the installation sequence in the box to
determine which load sensor is number 1, 2 and so on. You can use the IT functions from
either the front panel or Web interface. If you do not have the optional IT junction box,
some of the options described below will not appear on either interface.
Stability Test
The Stability test lets you check the A/D Raw count and filtered. With the IT option, it tests
and reports for each load cell. The test sends the load cell data to the analog-to-digital
converter and calculates the mean squared variation from the average reading, using 100
samples. The test passes if the mean squared variation is less than 5.0.
If the weighing system passes the stability test, the results show OK and the variation and
mean results are posted. FAIL indicates that the Mean Squared Variation is greater than
5.0 so the system is considered unstable. In that case, see the Troubleshooting Flow Charts
Section.
Summary of Contents for HI 6500 Series
Page 10: ...viii Contents ...
Page 20: ...10 Chapter 2 ...
Page 72: ...62 Chapter 4 ...
Page 96: ...86 Chapter 5 Profibus DP Commands and Parameters ...
Page 97: ...Instrument Configuration 87 Serial Option Termination ...
Page 126: ...116 Chapter 5 ...
Page 156: ...146 Chapter 8 ...
Page 163: ...Troubleshooting 153 General Troubleshooting Flow Chart Index ...
Page 164: ...154 Chapter 9 A Guideline Instability Electrical and Mechanical review ...
Page 165: ...Troubleshooting 155 A1 Checking for Unstable Components in a Weighing System ...
Page 166: ...156 Chapter 9 B Guidelines for Electrical Mechanical or Configuration Issues ...
Page 167: ...Troubleshooting 157 B1 Guidelines to Verify Electrical Installation ...
Page 168: ...158 Chapter 9 B2 Guidelines to Verify Mechanical Installation ...
Page 170: ...160 Chapter 9 C Integrated Technician and Stability Test Overview ...
Page 171: ...Troubleshooting 161 E Testing for Non Return to Zero System with IT Summing Card ...
Page 173: ...Troubleshooting 163 G Calibration Errors When Performing the Calibration ...
Page 174: ...164 Chapter 9 H Mechanical Installation ...
Page 175: ...Troubleshooting 165 J Electrical Inspection ...
Page 176: ...166 Chapter 9 K Installation Check Points ...
Page 177: ...Troubleshooting 167 M Weight Processor s Front Display is Blank or Locked ...
Page 178: ...168 Chapter 9 N Analog Out HI 6510 Only ...
Page 194: ...184 Appendix A Figure D1 Assembling the Swivel Mount Enclosure on the instrument pedestal ...
Page 196: ...186 Appendix A Dimensions ...
Page 200: ...4 HI 6500 Series USER GUIDE Index ...
Page 201: ......